GE OEC Medical Systems, Inc
384 N Wright Brothers Dr, Salt Lake City, Utah, United States
FEI 1720753
FDA has 6 inspections on file for GE OEC Medical Systems, Inc in Salt Lake City, Utah (FEI 1720753), from 2008 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2008–2023
- 0 OAI · 2 VAI · 4 NAI
- Inspections with 483 observations shown
- 2 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Rationale documented for no investigation
Records for complaints where no investigation was made do not always include the reason why.
MDR not evaluated per 820.198
A submitted MDR event was not evaluated in accordance with the requirements of the regulations regarding Failure Investigations.
Storage
Required records are not stored so as to minimize deterioration and prevent loss.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Show 2 more citations
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not implemented.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.180General requirements1 of 2
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1720753, one row per refused product line.
1 refusal on 2024-11-27 · charged REGISTERED on 1, NO REGISTR on 1 · no FDA sample analysis
FDA firm profile- 2024-11-27
C-ARM FLUOROSCOPIC X-RAY SYSTEM, MEDICAL DIAGNOSTIC X-RAY EQUIPMENT90RCC
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO REGISTR 536(a); Failure to file initial report
Division of Northeast Imports8RG-2463182-8/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1720753 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1720753. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.