FDA has 6 inspections on file for Medline Industries, LP in Arlington Heights, Illinois (FEI 1417411), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDA published 2 citations from this inspection. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on May 28, 2014. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Five regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2010–2026
2 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
↗
Requested under FOIA
4 requests, latest Jun 15, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesPosted 483 · 2 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.3.1

      Nonconforming product not identified and controlled

      Product which does not conform to product requirements was not adequately identified and controlled to prevent its unintended use or delivery.

    2. ISO 13485:2016 Clause 8.3.1

      Nonconforming product not adequately evaluated

      Nonconforming product was not adequately evaluated to determine the need for an investigation and notification of any external party responsible for the nonconformity.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Periodic equipment inspection lack of or inadequate procedu

      Procedures for conducting periodic inspections to ensure adherence to equipment maintenance schedules have not been established.

    3. Process changes - review, evaluation and revalidation

      A validated process was not reviewed and evaluated when changes or process deviations occurred.

    4. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    5. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.

    Show 1 more citation
    1. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    2. Process changes - review, evaluation and revalidation

      A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.

    3. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Periodic equipment inspection lack of or inadequate procedu

      Procedures for conducting periodic inspections to ensure adherence to equipment maintenance schedules have not been adequately established.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.

    2. Quantity manufactured

      The device history record does not include the quantity of devices manufactured.

    3. Analysis of data sources

      Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on Jun 15, 2026. Each one asked for a Form 483.

  • Inspection ending Feb 6, 2026VAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. May 19, 2026FDA-FOIA-2026-5147Form 483· 1 of 4 sitesRecords released
    2. Apr 17, 2026FDA-FOIA-2026-3890Form 483· 1 of 5 sitesOpen
  • Inspection date not given

    2 requests. FDA released records in answer to a request for its Form 483 on Jun 30, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 15, 2026FDA-FOIA-2026-5798Form 483Records released
    2. Feb 27, 2026FDA-FOIA-2026-2061Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Other FEIs recorded as Medline Industries, LP

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1417411 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1417411. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.