Hemagen Diagnostics Inc
2900 Loch Raven Rd, Baltimore, Maryland, United States
FEI 1181055
FDA has 6 inspections on file for Hemagen Diagnostics Inc in Baltimore, Maryland (FEI 1181055), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2010–2026
- 0 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Aug 21, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device have not been documented.
- ISO 13485:2016 Clause 7.4.1
Supplier performance not monitored
Supplier performance in meeting requirements for the purchased product was not monitored.
- ISO 13485:2016 Clause 4.2.5
Procedures for control of records not documented
Procedures to define the controls needed for the identification, storage, security, integrity, retrieval, retention time, and disposition of records have not been adequately documented.
- ISO 13485:2016 Clause 8.5.2
Corrective action not taken
Corrective action was not taken to eliminate the cause of nonconformities in order to prevent recurrence.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration procedures - content
Calibration procedures do not include specific directions and limits for accuracy and precision.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Aug 21, 2026. Each one asked for a Form 483.
Inspection ending May 14, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 14, 2026. FDAgent holds no copy that FDA posted.
- Aug 21, 2026FDA-FOIA-2026-8139Form 483· 1 of 4 sitesOpen
- Jul 1, 2026FDA-FOIA-2026-6260Form 483· 1 of 4 sitesRecords released
Inspection ending Jun 14, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Aug 22, 2024FDA-FOIA-2024-7440Form 483· 1 of 4 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.120Device labeling1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1181055 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1181055. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.