Neurotron, Inc.
12635 E Montview Blvd Ste 127, Aurora, Colorado, United States
FEI 1121806
FDA has 7 inspections on file for Neurotron, Inc. in Aurora, Colorado (FEI 1121806), from 2008 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 803.17.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2008–2026
- 0 OAI · 3 VAI · 4 NAI
- Inspections with 483 observations shown
- 3 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Oct 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.5.8
Product status identification was not maintained
Identification of product status was not maintained throughout production of product to ensure that only product that has passed the required inspections and tests or released under an authorized concession is dispatched, used, or installed.
- ISO 13485:2016 Clause 7.4.1
Procedures for purchasing process not documented
Procedures to ensure that purchased product conforms to specified purchasing information have not been adequately documented.
- ISO 13485:2016 Clause 7.5.4
Servicing procedures and reference materials and measurements not documented
The servicing procedures for performing servicing activities and verifying that product requirements are met were not adequately documented.
- ISO 13485:2016 Clause 7.4.3
Purchased product verification activities not established or implemented
The inspection or other activities necessary for ensuring that purchased product meets specified purchasing requirements were not adequately established and implemented.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for planning and documenting action needed and implementing such action and reviewing the effectiveness of corrective action taken have not been adequately documented.
Show 1 more citation
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2026. Each one asked for a Form 483.
Inspection ending Aug 21, 2026VAI
2 requests. In the inspection history above.
- Oct 16, 2026FDA-FOIA-2026-10081Form 483· 1 of 5 sitesOpen
- Oct 6, 2026FDA-FOIA-2026-9717Form 483· 1 of 5 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 820.200Servicing1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1121806 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1121806. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.