Skip to content

Neurotron, Inc.

12635 E Montview Blvd Ste 127, Aurora, Colorado, United States

FEI 1121806

FDA has 7 inspections on file for Neurotron, Inc. in Aurora, Colorado (FEI 1121806), from 2008 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 803.17.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2008–2026
0 OAI · 3 VAI · 4 NAI
Inspections with 483 observations shown
3 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Oct 16, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.5.8

      Product status identification was not maintained

      Identification of product status was not maintained throughout production of product to ensure that only product that has passed the required inspections and tests or released under an authorized concession is dispatched, used, or installed.

    2. ISO 13485:2016 Clause 7.4.1

      Procedures for purchasing process not documented

      Procedures to ensure that purchased product conforms to specified purchasing information have not been adequately documented.

    3. ISO 13485:2016 Clause 7.5.4

      Servicing procedures and reference materials and measurements not documented

      The servicing procedures for performing servicing activities and verifying that product requirements are met were not adequately documented.

    4. ISO 13485:2016 Clause 7.4.3

      Purchased product verification activities not established or implemented

      The inspection or other activities necessary for ensuring that purchased product meets specified purchasing requirements were not adequately established and implemented.

    5. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for planning and documenting action needed and implementing such action and reviewing the effectiveness of corrective action taken have not been adequately documented.

    Show 1 more citation
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities have not been adequately established.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  7. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 21, 2026VAI

    2 requests. In the inspection history above.

    1. Oct 16, 2026FDA-FOIA-2026-10081Form 483· 1 of 5 sitesOpen
    2. Oct 6, 2026FDA-FOIA-2026-9717Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 7 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1121806 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1121806. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.