Pedicraft, Inc.
4134 Saint Augustine Rd, Jacksonville, Florida, United States
FEI 1043568
FDA has 5 inspections on file for Pedicraft, Inc. in Jacksonville, Florida (FEI 1043568), from 2011 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2011–2025
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Sep 16, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed and is incomplete.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file was not established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Show 1 more citation
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Show 3 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 16, 2025. It asked for a Form 483.
Inspection ending Jul 25, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 1, 2025. FDAgent holds no copy that FDA posted.
- Sep 16, 2025FDA-FOIA-2025-10947Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit4 of 4
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.184Device history record1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1043568 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1043568. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.