Nipro Renal Solutions USA, Corporation

509 Fishing Creek Rd, Lewisberry, Pennsylvania, United States

FEI 1000142899

FDA has 4 inspections on file for Nipro Renal Solutions USA, Corporation in Lewisberry, Pennsylvania (FEI 1000142899), from 2013 to 2026. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDA published 15 citations from this inspection. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
4, 2013–2026
1 OAI · 1 VAI · 2 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
↗
Requested under FOIA
3 requests, latest Sep 14, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 15 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.9

      Procedures to control design and development changes not documented

      Procedures to control design and development changes were not adequately documented.

    2. ISO 13485:2016 Clause 6.3

      Maintenance records not maintained

      Maintenance records have not been adequately maintained.

    3. ISO 13485:2016 Clause 8.2.2

      Complaint handling records not maintained

      Complaint handling records were not adequately maintained.

    4. ISO 13485:2016 Clause 4.2.3

      Medical device file not established and maintained

      Medical device files have not been adequately established and maintained.

    5. ISO 13485:2016 Clause 7.5.1

      Medical device or batch records not adequate

      A record for each medical device or batch of medical devices that provides specified traceability and identifies the amount manufactured and amount approved for distribution has not been established and maintained.

    Show 10 more citations
    1. ISO 13485:2016 Clause 8.2.4

      Internal audits not conducted to ensure QMS meets requirements and is effective

      Internal audits have not been adequately conducted at planned intervals to determine whether the QMS is effectively implemented and maintained.

    2. ISO 13485:2016 Clause 6.4.1

      Work environment requirements and procedures not documented

      Requirements for the work environment, and the procedures to monitor and control the work environment, were not documented where conditions for the work environment can have an adverse effect on product quality.

    3. ISO 13485:2016 Clause 8.5.2

      Corrective action not taken

      Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.

    4. ISO 13485:2016 Clause 8.3.1

      Nonconforming product not identified and controlled

      Product which does not conform to product requirements was not identified and controlled to prevent its unintended use or delivery.

    5. ISO 13485:2016 Clause 7.3.10

      Design and development files not maintained

      A design and development file was not maintained for each medical device type or medical device family.

    6. ISO 13485:2016 Clause 7.4.1

      Supplier performance not monitored

      Supplier performance in meeting requirements for the purchased product was not monitored.

    7. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.

    8. ISO 13485:2016 Clause 8.3.1

      Records of nonconforming product and actions taken not maintained

      Records of the nature of nonconforming product and any subsequent action taken, including the evaluation, and any investigation and the rationale for decisions, were not maintained.

    9. ISO 13485:2016 Clause 7.5.6

      Records of validation of processes and necessary actions not maintained

      Records of the results and conclusion of validation of processes and necessary actions from the validation were not maintained.

    10. ISO 13485:2016 Clause 8.2.6

      Evidence of conformity to the acceptance criteria not maintained

      Evidence of conformity to the acceptance criteria for product monitoring and measurement was not adequately maintained.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 14, 2026. Each one asked for a Form 483.

  • Inspection ending Mar 27, 2026OAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. Sep 14, 2026FDA-FOIA-2026-8875Form 483 + EIROpen
    2. May 13, 2026FDA-FOIA-2026-4917Form 483· 1 of 5 sitesRecords released
  • Inspection ending Mar 3, 2026

    1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Aug 20, 2026. FDAgent holds no copy that FDA posted.

    1. Aug 13, 2026FDA-FOIA-2026-7837Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000142899 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000142899. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.