Nipro Renal Soultions USA, Corporation Form 483, March 2026

FDA issued this Form 483 to Nipro Renal Soultions USA, Corporation in Lewisberry, Pennsylvania (FEI 1000142899) at the end of a device inspection that ended March 27, 2026. FDA classified the inspection Official Action Indicated (OAI) and issued the facility a warning letter on July 16, 2026.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 27, 2026
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Nov 17, 2026
235 days after the inspection
Facility
FEI 1000142899
Lewisberry, Pennsylvania

FDA's 15 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 7.3.9

    Procedures to control design and development changes not documented

    Procedures to control design and development changes were not adequately documented.

  2. ISO 13485:2016 Clause 6.3

    Maintenance records not maintained

    Maintenance records have not been adequately maintained.

  3. ISO 13485:2016 Clause 8.2.2

    Complaint handling records not maintained

    Complaint handling records were not adequately maintained.

  4. ISO 13485:2016 Clause 4.2.3

    Medical device file not established and maintained

    Medical device files have not been adequately established and maintained.

  5. ISO 13485:2016 Clause 7.5.1

    Medical device or batch records not adequate

    A record for each medical device or batch of medical devices that provides specified traceability and identifies the amount manufactured and amount approved for distribution has not been established and maintained.

  6. ISO 13485:2016 Clause 8.2.4

    Internal audits not conducted to ensure QMS meets requirements and is effective

    Internal audits have not been adequately conducted at planned intervals to determine whether the QMS is effectively implemented and maintained.

  7. ISO 13485:2016 Clause 6.4.1

    Work environment requirements and procedures not documented

    Requirements for the work environment, and the procedures to monitor and control the work environment, were not documented where conditions for the work environment can have an adverse effect on product quality.

  8. ISO 13485:2016 Clause 8.5.2

    Corrective action not taken

    Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.

  9. ISO 13485:2016 Clause 8.3.1

    Nonconforming product not identified and controlled

    Product which does not conform to product requirements was not identified and controlled to prevent its unintended use or delivery.

  10. ISO 13485:2016 Clause 7.3.10

    Design and development files not maintained

    A design and development file was not maintained for each medical device type or medical device family.

  11. ISO 13485:2016 Clause 7.4.1

    Supplier performance not monitored

    Supplier performance in meeting requirements for the purchased product was not monitored.

  12. ISO 13485:2016 Clause 4.1.2

    Risk-based approach not applied to QMS processes

    A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.

  13. ISO 13485:2016 Clause 8.3.1

    Records of nonconforming product and actions taken not maintained

    Records of the nature of nonconforming product and any subsequent action taken, including the evaluation, and any investigation and the rationale for decisions, were not maintained.

  14. ISO 13485:2016 Clause 7.5.6

    Records of validation of processes and necessary actions not maintained

    Records of the results and conclusion of validation of processes and necessary actions from the validation were not maintained.

  15. ISO 13485:2016 Clause 8.2.6

    Evidence of conformity to the acceptance criteria not maintained

    Evidence of conformity to the acceptance criteria for product monitoring and measurement was not adequately maintained.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.