Skip to content

Hologic, Inc

250 Campus Dr, Marlborough, Massachusetts, United States

FEI 1000120743

FDA has 6 inspections on file for Hologic, Inc in Marlborough, Massachusetts (FEI 1000120743), from 2011 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the seven FOIA requests for this site has asked FDA for this inspection's 483.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.100(a).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
6, 2011–2025
2 OAI · 0 VAI · 4 NAI
Inspections with 483 observations shown
2 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
7 requests, latest Jan 8, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    5. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

  3. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
  4. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Seven requests to FDA under the Freedom of Information Act named this site, most recently on Jan 8, 2026.

  • Inspection ending Nov 4, 2025OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 11, 2026. FDAgent holds no copy that FDA posted.

    1. Jan 8, 2026FDA-FOIA-2026-267Form 483· 1 of 5 sitesRecords released
  • Inspection ending Sep 24, 2024OAI

    5 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 6, 2025. FDAgent holds no copy that FDA posted.

    1. Oct 6, 2025FDA-FOIA-2025-11607EIROpen
    2. Aug 28, 2025FDA-FOIA-2025-10344Form 483Records released
    3. Jun 23, 2025FDA-FOIA-2025-7933Form 483Records released
    4. Mar 4, 2025FDA-FOIA-2025-1962Form 483· 1 of 3 sitesClosed
    5. Nov 8, 2024FDA-FOIA-2024-9926Form 483· 1 of 2 sitesClosed
  • Inspection ending Jun 29, 2017

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Oct 14, 2021FDA-FOIA-2021-6994Inspection recordsClosed over fees

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000120743 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000120743. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.