Skip to content

Metrex Research, LLC.

28210 Wick Rd, Romulus, Michigan, United States

FEI 1000117344DUNS 145963778

FDA has 5 inspections on file for Metrex Research, LLC. in Romulus, Michigan (FEI 1000117344), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued. Two of the five FOIA requests for this site have asked FDA for this inspection's 483.

1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing and labeling, through Dec 31, 2027. Its registrant is Metrex Research, LLC (D-U-N-S 145963778).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2010–2026
0 OAI · 2 VAI · 3 NAI
Inspections with 483 observations shown
2 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
5 requests, latest FY2023
Warning letters
None on file
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Microbiological testing

      Each batch of drug product required to be free of objectionable microorganisms is not tested through appropriate laboratory testing.

    2. Label storage access limited to authorized personnel

      Access to the storage area for labels and labeling materials is not limited to authorized personnel.

    3. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, sampling plans and test procedures designed to assure that components conform to appropriate standards of identity, strength, quality and purity.

    4. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not cleaned and maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    5. Computer control of master formula records

      Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.

    Show 2 more citations
    1. Establish reliability of supplier's C of A

      Establishment of the reliability of the component supplier's report of analyses is deficient in that the test results are not appropriately validated at appropriate intervals.

    2. Calibration/Inspection/Checking not done

      Routine calibration, inspection and checking of automatic and mechanical equipment is not performed according to a written program designed to assure proper performance.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    2. Documentation

      Process validation activities have not been adequately approved.

  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Five requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.

  • Inspection ending May 1, 2026VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 22, 2026. FDAgent holds no copy that FDA posted.

    1. Jul 16, 2026FDA-FOIA-2026-6786Form 483Records released
    2. Jul 8, 2026FDA-FOIA-2026-6512Form 483· 1 of 5 sitesRecords released
  • Inspection ending May 20, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-93Form 483Closed
  • Inspection ending May 5, 2017

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-4401Inspection recordsClosed
  • Inspection ending Jul 15, 2015

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Apr 12, 2022FDA-FOIA-2022-2669Inspection recordsClosed over fees

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 5 inspections, with how many of them cited each.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000117344, one row per refused product line.

1 refusal on 2026-09-15 · charged NOT LISTED on 1 · no FDA sample analysis

FDA firm profile
  • 2026-09-15

    DISINFECTANT, MEDICAL DEVICES80LRJ

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Northeast Imports823-0226399-9/10001/1

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, packing, labeling
Registration runs through
Dec 31, 2027
Registrant
Metrex Research, LLCD-U-N-S 145963778

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000117344 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000117344. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.