Skip to content

21 CFR 211.134: Drug product inspection

FDA's inspection data lists 74 Form 483 citations of 21 CFR 211.134 at 71 facilities, from May 2009 to November 14, 2025.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
74
At 72 inspections
Facilities cited
71
Distinct FEIs
Most recent citation
Nov 14, 2025
Tenco Assemblies, Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 211.134 from the Electronic Code of Federal Regulations.

§ 211.134 Drug product inspection.

(a) Packaged and labeled products shall be examined during finishing operations to provide assurance that containers and packages in the lot have the correct label.

(b) A representative sample of units shall be collected at the completion of finishing operations and shall be visually examined for correct labeling.

(c) Results of these examinations shall be recorded in the batch production or control records.

21 CFR Part 211: Current Good Manufacturing Practice for Finished Pharmaceuticals, Subpart GRead on eCFR

Citations by year

Form 483 citations of 21 CFR 211.134 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 16 in 2011

Form 483 citations of 21 CFR 211.134 by inspection year
YearForm 483 citations
20098
20106
201116
20123
20135
20149
20151
20162
20177
20183
20198
20201
20221
20231
20241
20252

How FDA words this citation

FDA's short descriptions and citation text on these 74 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Correct labels during finishing operations29 citations
  • Examinations documented28 citations
  • Representative samples after completion17 citations

Citation text

  • “Packaged and labeled products are not examined during finishing operations to provide assurance that containers and packages in the lot have the correct label.”29 citations
  • “The results of the examination of the packaged and labeled products were not documented in the batch production or control records.”28 citations
  • “Samples of representative units were not visually examined for correct labeling at the completion of finishing operations.”9 citations
  • “Samples of representative units were not collected and visually examined for correct labeling at the completion of finishing operations.”6 citations
Show 1 more
  • “Samples of representative units were not collected for correct labeling at the completion of finishing operations.”2 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 211.134, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 211.134 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 2 observations tie this way. The highlighted sentence is the matched wording.

  • LeeSar Inc2026-09-23 · Observation 9

    Samples of representative units were not visually examined for correct labeling at the completion of finishing operations. Specifically, you failed to visually inspect representative samples of your finished drug products to ensure they contain the correct labeling. Your Visual Inspection Program requires 100% visual inspection of your syringe and IV bag drug products for defects such as leaks, particles, metal, fiber and hair. Inspectors …

    This observation opens with FDA's standard wording for 21 CFR 211.134(b).

  • SCA Pharmaceuticals, Inc.2019-11-07 · Observation 4

    The results of the examination of the packaged and labeled products were not documented in the batch production or control records. Specifically, I observed during the visual inspection by employees rejects from the 100% and the AQL were placed in the same bin, thus not allowing for accurate data to be entered in the batch records to determine if both passed your specification for visual inspection as defined in SOP COM-013-LR. For the AQL, …

    This observation opens with FDA's standard wording for 21 CFR 211.134(c).

Often cited at the same inspection

Of the 72 inspections where FDA cited 21 CFR 211.134, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 211.134

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 211.134; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.