St. Jude Medical Inc. Form 483, February 2017
FDA issued this Form 483 to St. Jude Medical Inc. in Sylmar, California (FEI 2017865) at the end of a device inspection that ended February 17, 2017. FDA classified the inspection Official Action Indicated (OAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Feb 17, 2017
- Devices · Compliance: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Jan 17, 2024
- 2,525 days after the inspection
FDA's 6 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
After this inspection
FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.
- Warning LetterApr 12, 2017CDRH · 54 days after this inspection ended
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 2017865, newest first.
- Form 483, July 2015FDA record date Jul 10, 2015 · Posted Jan 17, 2024
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.