Skip to content

Sheffield Precision Medical Limited Form 483, October 2025

FDA issued this Form 483 to Sheffield Precision Medical Limited in Sheffield, United Kingdom (FEI 3009871135) at the end of a device inspection that ended October 23, 2025. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 12, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Oct 23, 2025
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Nov 12, 2026
385 days after the inspection
Facility
FEI 3009871135
Sheffield, United Kingdom

FDA's 6 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Adjustment

    Inherent limitations or allowable tolerances are not posted on or near and are not readily available to personnel performing adjustments of equipment requiring periodic adjustments.

  2. Maintenance schedule, Lack of or inadequate schedule

    Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.

  3. Lack of or inadequate In-process acceptance procedures

    Procedures for the acceptance and control of in-process product have not been established.

  4. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

  5. DHR content

    The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

  6. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.