Skip to content

Fujifilm Medical Systems U.S.A., Inc. Form 483, April 2015

FDA issued this Form 483 to Fujifilm Medical Systems U.S.A., Inc. in Wayne, New Jersey (FEI 2431293) at the end of a device inspection that ended April 9, 2015. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Apr 9, 2015
Devices · Postmarket Assurance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Jan 17, 2024
3,205 days after the inspection
Facility
FEI 2431293
Wayne, New Jersey

FDA's 5 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  2. Report of risk to health

    A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

  3. Investigation of device failures

    Complaints involving the possible failure of a device and labeling to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

  4. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  5. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterAug 12, 2015CDRH · 125 days after this inspection ended

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.