Skip to content

Tacoma-Pierce County Blood Bank

220 S I St, Tacoma, Washington, United States

FEI 3075335

FDA has 8 inspections on file for Tacoma-Pierce County Blood Bank in Tacoma, Washington (FEI 3075335), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection: 1 under 21 CFR Part 211, the manufacturing rules for finished drugs, and 1 under 21 CFR Part 606, the manufacturing rules for blood and blood components. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 606.100(b) and 21 CFR 606.100(c).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
8, 2009–2023
0 OAI · 3 VAI · 5 NAI
Inspections with 483 observations shown
4 of 8
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written calibration / inspection records not kept

      Records of the of automatic, mechanical or electronic equipment, including computers or related systems are not maintained.

    2. Equipment observed, standardized, calibrated

      Equipment used in the collection, processing and storage and distribution of blood and blood components is not calibrated on a regularly scheduled basis as prescribed in the SOP Manual.

  2. NAINo objectionable conditions that need follow-upBiologics · Vaccines and Allergenic Products6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the processing of blood and blood components for allogeneic transfusion were not always followed.

    2. Provide space for quarantine of unsuitable for use items

      Failure to provide adequate space for the quarantine and disposition of products and reagents not suitable for use.

    3. Thorough investigations

      Failure to perform a thorough investigation of a failure of a lot or unit to meet any of its specifications.

    4. Equipment calibration frequency

      Equipment used in the processing of blood and blood components is not observed, standardized and calibrated with at least the frequency required.

    5. CPR certification

      Failure to assure that an individual who is currently certified in cardiopulmonary resuscitation is located on the premises whenever collections of blood or blood components are performed.

    Show 1 more citation
    1. Required records

      Failure to maintain general records.

  3. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  4. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  5. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Following manufacturer's instructions

      Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.

    2. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection and storage of blood and blood components for homologous transfusion and further manufacturing purposes are not always followed.

    3. Thorough investigations

      Failure to perform a thorough investigation of an unexplained discrepancy.

  7. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection of blood and blood components for homologous transfusion are not always followed.

  8. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 8 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3075335 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3075335. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.