Skip to content

Seniorlife Technologies Inc

1400 112th Ave Se Ste 100, Bellevue, Washington, United States

FEI 3027535342

FDA has one inspection on file for Seniorlife Technologies Inc in Bellevue, Washington (FEI 3027535342). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH). FDA has also posted a Form 483 dated Feb 20, 2026 that FDAgent has not matched to an inspection on the Data Dashboard.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 7 citations from this inspection: 6 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483, and two more for the 483 from a later inspection, ending Feb 20, 2026, that FDA's Data Dashboard doesn't list yet.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2025
Inspections with 483 observations shown
1 of 1
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Requested under FOIA
3 requests, latest May 19, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    4. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

    5. Design control - no procedures

      Procedures for design control have not been established.

    Show 2 more citations
    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on May 19, 2026. Each one asked for a Form 483.

  • Inspection ending Feb 20, 2026

    2 requests. Not listed in FDA’s Data Dashboard yet. FDA has posted its Form 483.

    1. May 19, 2026FDA-FOIA-2026-5142Form 483· 1 of 4 sitesRecords released
    2. Apr 13, 2026FDA-FOIA-2026-3674Form 483· 1 of 5 sitesRecords released
  • Inspection ending Mar 26, 2025OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Nov 19, 2025. FDAgent holds no copy that FDA posted.

    1. Sep 4, 2025FDA-FOIA-2025-10563Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3027535342 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3027535342. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.