TRUE DIAGNOSTICS, INC
5919 Farnsworth Ct, Carlsbad, California, United States
FEI 3027500452
FDA has one inspection on file for TRUE DIAGNOSTICS, INC in Carlsbad, California (FEI 3027500452). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483, and one more for the 483 from a later inspection, ending Dec 29, 2025, that FDA's Data Dashboard doesn't list yet.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2022
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Sep 3, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate In-process acceptance procedures
Procedures for the control of in-process product have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Show 1 more citation
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 3, 2026. Each one asked for a Form 483.
Inspection ending Dec 29, 2025
1 request. Not listed in FDA’s Data Dashboard yet.
- Aug 20, 2026FDA-FOIA-2026-8079Form 483· 1 of 5 sitesOpen
Inspection ending Mar 9, 2022VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 2, 2026. FDAgent holds no copy that FDA posted.
- Sep 3, 2026FDA-FOIA-2026-8588Form 483· 1 of 5 sitesRecords released
- Jul 28, 2025FDA-FOIA-2025-9161Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.184Device history record1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3027500452 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3027500452. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.