iRhythm Technologies, Inc.
6550 Katella Ave Ste 200, Cypress, California, United States
FEI 3021769057
FDA has 3 inspections on file for iRhythm Technologies, Inc. in Cypress, California (FEI 3021769057), from 2013 to 2022. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH). FDA has also posted a Form 483 dated Jul 31, 2024 that FDAgent has not matched to an inspection on the Data Dashboard.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. Six of the nine FOIA requests for this site asked FDA for records from two later inspections that FDA's Data Dashboard doesn't list yet, the latest ending Jul 31, 2024.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2013–2022
- 1 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 3
- 1 posted 483 not matched to an inspection
- Requested under FOIA
- 9 requests, latest May 28, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 4 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Nine requests to FDA under the Freedom of Information Act named this site, most recently on May 28, 2025.
Inspection ending Jul 31, 2024
4 requests. Not listed in FDA’s Data Dashboard yet. FDA has posted its Form 483.
- Oct 16, 2024FDA-FOIA-2024-9148EIR· 1 of 2 sitesReleased in part
- Aug 20, 2024FDA-FOIA-2024-7405Form 483· 1 of 2 sitesClosed
- Aug 8, 2024FDA-FOIA-2024-7105Form 483Closed
- Aug 2, 2024FDA-FOIA-2024-6895Form 483Closed
Inspection ending Jul 15, 2024
2 requests. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Aug 16, 2024FDA-FOIA-2024-7313Form 483· 1 of 2 sitesClosed
- Aug 14, 2024FDA-FOIA-2024-7258Form 483Closed
Inspection date not given
3 requests. FDA released records in answer to a request for its Form 483 on Oct 3, 2025. FDAgent holds no copy that FDA posted.
- May 28, 2025FDA-FOIA-2025-7076Form 483Records released
- Aug 7, 2024FDA-FOIA-2024-7060Form 483· 1 of 2 sitesClosed
- Aug 6, 2024FDA-FOIA-2024-6991Form 483· 1 of 2 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3021769057 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3021769057. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.