RetroPerc, Inc.
111 Town Square Pl Ste 1203, Jersey City, New Jersey, United States
FEI 3019946783
FDA has one inspection on file for RetroPerc, Inc. in Jersey City, New Jersey (FEI 3019946783). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection: 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2024
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Apr 30, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 1 more citation
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Apr 30, 2025. Each one asked for a Form 483.
Inspection ending Dec 18, 2024VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 30, 2025FDA-FOIA-2025-3894Form 483· 1 of 4 sitesClosed
- Feb 19, 2025FDA-FOIA-2025-1534Form 483· 1 of 4 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3019946783 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3019946783. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.