Intelligent Implant Systems
3300 International Airport Dr Ste 1100, Charlotte, North Carolina, United States
FEI 3011121367
FDA has one inspection on file for Intelligent Implant Systems in Charlotte, North Carolina (FEI 3011121367). It ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 1, in 2018
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of MDR Investigation
Investigation records of MDR reportable complaints do not include required information.
Design validation - production units
The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Design verification - documentation
The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not adequately documented in the design history file.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
Show 2 more citations
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Supplier notification of changes
There is no agreement with suppliers and contractors to notify you of changes in the product or service.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.120Device labeling1 of 1
- 21 CFR 820.198Complaint files1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011121367 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011121367. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.