FDA has 2 inspections on file for Sonesta Medical AB in Solna, Sweden (FEI 3010496070), from 2017 to 2023. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 7 citations from this inspection: 6 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.30(c) and 21 CFR 820.50.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2017–2023
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jan 11, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been maintained.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    4. Lack of procedures, or not maintained

      Document control procedures have not been established and maintained. Electronic signatures based on non-biometrics are used, but they do not meet the id/password uniqueness requirements of 21 CFR Part 11.

    5. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    Show 2 more citations
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    4. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 2 more citations
    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jan 11, 2024. It asked for a Form 483.

  • Inspection ending Jun 7, 2023OAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jan 11, 2024FDA-FOIA-2024-381Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010496070 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010496070. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.