Skip to content

Surgical Theater Inc

23645 Mercantile Rd Ste M, Beachwood, Ohio, United States

FEI 3010197287

FDA has 5 inspections on file for Surgical Theater Inc in Beachwood, Ohio (FEI 3010197287), from 2014 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.100(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2014–2026
0 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Aug 21, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.5

      Records of design and development reviews not maintained

      Records of the results of the design and development reviews and any necessary actions were not adequately maintained or include identification of the design under review, the participants involved, and the date of the review.

    2. ISO 13485:2016 Clause 4.2.4

      Document Control Procedure Inadequate

      The document control procedure does not adequately define controls needed to review and approve documents for adequacy before use.

    3. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.

    2. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    3. Management representative's reports

      The management representative has not fully reported on the quality system performance to management with executive responsibility.

    4. Design review - documentation

      The design review results, including identification of the design, the date and the individual(s) performing the review, were not documented in the design history file.

    5. Product rework adverse effects {see also 820.184}

      Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Design validation - software validation not performed

      Validation of device software is incomplete.

    3. Document review, approval documentation

      The documentation of approval of documents does not include the document approval date and the signature of the approving official.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Design input - documentation

      Design input requirements were not adequately documented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Aug 21, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 11, 2026VAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 20, 2026. FDAgent holds no copy that FDA posted.

    1. Aug 21, 2026FDA-FOIA-2026-8135Form 483· 1 of 4 sitesRecords released
    2. Aug 4, 2026FDA-FOIA-2026-7501Form 483· 1 of 5 sitesRecords released
    3. Aug 3, 2026FDA-FOIA-2026-7441Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010197287 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010197287. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.