Skip to content

Alan Neuromedical Technologies

2002 Timberloch Pl Ste 200, The Woodlands, Texas, United States

FEI 3010141624

FDA has 3 inspections on file for Alan Neuromedical Technologies in The Woodlands, Texas (FEI 3010141624), from 2011 to 2018. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Nov 5, 2014. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2011–2018
2 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Complaints

      Complaint files are not adequately maintained.

    3. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sponsors' general responsibilities

      For an investigational study, proper monitoring was not ensured.

    2. Availability of records of non-significant risk studies

      For an investigation subject to the abbreviated requirements, records are not consolidated in one location and available for FDA inspection and copying.

    3. Consent form not approved/signed/dated

      Informed consent was not properly documented in that the written informed consent used in the study was not signed by the subject or the subject?s legally authorized representative at the time of consent and was not dated by the subject or the subject's legally authorized representative at the time of consent.

    4. Sponsor correspondence records inadequate

      Records relating to correspondence with a monitor, an investigator, and an IRB, including required reports are not all accurate, complete, and current.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigator lack of informed consent

      Informed consent was not obtained in accordance with the regulations regarding the protection of human subjects.

    2. No investigation agreement statement of commitment

      A signed agreement that includes a statement of the investigator's commitment to conduct an investigation in accordance with the agreement, investigational plan, applicable FDA regulations, conditions of approval imposed by the reviewing IRB, and conditions of approval imposed by the FDA was not obtained from each participating investigator.

    3. Investigator device accountability inadequate

      Records of receipt, use, and disposal of a device that relate to the type and quantity, dates of receipt, and batch number or code mark of the device are not all accurate, complete, and current.

    4. Investigator records of exposure to device inadequate

      Records of each subject's exposure to the device, including the date and time of each use and the use of any other therapy are not all complete and current.

    5. Sponsor began study before IRB/FDA approval

      Part of an investigation was initiated before FDA approval and IRB approval.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 9

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010141624 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010141624. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.