Medspira, LLC
2718 Summer St NE, Minneapolis, Minnesota, United States
FEI 3008414621
FDA has 4 inspections on file for Medspira, LLC in Minneapolis, Minnesota (FEI 3008414621), from 2014 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's records.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2014–2025
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Oct 16, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Show 1 more citation
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Show 3 more citations
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2025.
Inspection ending Jun 27, 2025NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its EIR on May 11, 2026. FDAgent holds no copy that FDA posted.
- Oct 16, 2025FDA-FOIA-2025-12078EIRRecords released
- Jul 21, 2025FDA-FOIA-2025-8889Form 483· 1 of 5 sitesClosed
Inspection ending Jan 16, 2024VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Sep 12, 2024FDA-FOIA-2024-8136Form 483· 1 of 5 sitesClosed
- Jul 30, 2024FDA-FOIA-2024-6621Form 483· 1 of 4 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.40[Reserved]1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008414621 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008414621. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.