Skip to content

Product Design and Development Corp.

16337 Westwoods Business Park, Ellisville, Missouri, United States

FEI 3008146975

FDA has 4 inspections on file for Product Design and Development Corp. in Ellisville, Missouri (FEI 3008146975), from 2011 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 3 citations from this inspection: 2 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.184 and 21 CFR 820.25(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2011–2025
0 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Apr 30, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training records

      Personnel training is not documented.

    2. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    3. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been established.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    3. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been established.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    5. Design review - Lack of or inadequate procedures

      Procedures for design review have not been established.

    Show 7 more citations
    1. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been established.

    2. Design input - Lack of or inadequate procedures

      Procedures for design input have not been established.

    3. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.

    4. Quality Audit/Reaudit - conducted

      Quality audits and reaudits have not been performed.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    6. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    7. Training records

      Personnel training is not documented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Apr 30, 2025. Each one asked for a Form 483.

  • Inspection ending Jan 24, 2025NAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Apr 30, 2025FDA-FOIA-2025-3911Form 483Closed
    2. Feb 14, 2025FDA-FOIA-2025-1459Form 483· 1 of 2 sitesClosed
  • Inspection ending Mar 17, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Apr 28, 2022FDA-FOIA-2022-3077Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008146975 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008146975. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.