Skip to content

Hyperbaric America LLC

1902 7th Ct N, Lake Worth, Florida, United States

FEI 3007139001

FDA has 2 inspections on file for Hyperbaric America LLC in Lake Worth, Florida (FEI 3007139001), all in 2009. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 12 citations from this inspection: 11 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2009
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR QA procedures and specifications

      The device master record does not include or refer to the location of quality assurance procedures and specifications.

    2. Design verification - Lack of or inadequate procedures

      Design verification did not confirm that the design output meets the design input requirements.

    3. Design history file

      The design history file was not established or maintained.

    4. Acceptance records

      The device history record does not include acceptance records that demonstrate the device is manufactured in accordance with the device master record.

    5. Design input procedures - appropriateness

      Procedures to ensure that a device's design input requirements are appropriate and address its intended use, including user/patient needs, were not established, defined, documented, complete, and implemented.

    Show 7 more citations
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    2. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not established, defined, documented, and implemented.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not established, defined, and documented.

    5. Design control - no procedures

      Procedures to control the design process of the device were not established, defined, documented, and implemented.

    6. Quality Audit/Reaudit - conducted

      Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.

    7. Management review - Lack of or inadequate procedures

      Procedures for management review are not documented, complete, and implemented.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007139001 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007139001. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.