Hyperbaric America LLC
1902 7th Ct N, Lake Worth, Florida, United States
FEI 3007139001
FDA has 2 inspections on file for Hyperbaric America LLC in Lake Worth, Florida (FEI 3007139001), all in 2009. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 12 citations from this inspection: 11 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2009
- 1 OAI · 0 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR QA procedures and specifications
The device master record does not include or refer to the location of quality assurance procedures and specifications.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Design history file
The design history file was not established or maintained.
Acceptance records
The device history record does not include acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Design input procedures - appropriateness
Procedures to ensure that a device's design input requirements are appropriate and address its intended use, including user/patient needs, were not established, defined, documented, complete, and implemented.
Show 7 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not established, defined, documented, and implemented.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not established, defined, and documented.
Design control - no procedures
Procedures to control the design process of the device were not established, defined, documented, and implemented.
Quality Audit/Reaudit - conducted
Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.
Management review - Lack of or inadequate procedures
Procedures for management review are not documented, complete, and implemented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.181Device master record1 of 1
- 21 CFR 820.184Device history record1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007139001 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007139001. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.