Degania Medical Devices Pvt. Ltd.
Plot No 251, Imt Manesar, Gurgaon, India
FEI 3005483737
FDA has 4 inspections on file for Degania Medical Devices Pvt. Ltd. in Gurgaon, India (FEI 3005483737), from 2010 to 2019. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.75(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 4, 2010–2019
- 1 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities have not been adequately documented .
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Document review, approval by designated individual
Documents were not reviewed by designated individual(s) prior to issuance .
Documentation
Process validation activities have not been adequately documented.
Periodic equipment inspections
Periodic inspections of equipment conducted to ensure adherence to applicable maintenance schedules were not documented.
Equipment control activity documentation
Equipment calibrations have not been documented.
Document locations, Dissemination, etc.
Documents were not available at all locations for which they are designated, used, or otherwise necessary.
Show 2 more citations
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Implementing Personnel Procedures, Health, Cleanliness.
Requirements addressing the clothing of employees have not been implemented.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005483737, one row per refused product line.
1 refusal on 2019-01-31 · charged NO 510(K) on 1 · no FDA sample analysis
FDA firm profile- 2019-01-31
WOUND DRAIN CATHETER SYSTEM79OTK
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northeast Imports923-4418972-2/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005483737 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005483737. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.