Spectranetics Corporation
5055 Brandin Ct, Fremont, California, United States
FEI 3005462046
FDA has 5 inspections on file for Spectranetics Corporation in Fremont, California (FEI 3005462046), from 2009 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2009–2016
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - documentation
Design input requirements were not adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Patient Weight
An individual medical device manufacturer report submitted per FDA Form 3500A did not contain in Block A the patient's weight.
Patient Age or Date of Birth
An individual medical device manufacturer report submitted per FDA Form 3500A did not contain in Block A the patient's age or date of birth.
Patient Name or Other Identifier
An individual medical device manufacturer report submitted per FDA Form 3500A did not contain in Block A the patient name or other identifier.
Patient Gender
An individual medical device manufacturer report submitted per FDA Form 3500A did not contain in Block A the patient's gender.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Nature and details of complaint
Records of complaint investigations do not include the nature and details of the complaint.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not complete.
Sampling plans
Sampling plans are not established and based on valid statistical rationale.
Show 1 more citation
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product throughout all stages of receipt, production, packaging, distribution, and installation were not complete.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 4
- 21 CFR 820.30Design controls1 of 4
- 21 CFR 820.60Identification1 of 4
- 21 CFR 820.198Complaint files1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Other FEIs recorded as Spectranetics Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Spectranetics CorporationFEI 3009784280Fremont, California, United States
Last inspected 2020-09-17NAI
Spectranetics CorporationFEI 3007284006Colorado Springs, Colorado, United States
Last inspected 2019-03-01NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005462046 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005462046. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.