Skip to content

Synthes, Inc.

1302 Wrights Ln E, West Chester, Pennsylvania, United States

FEI 3005180112

FDA has 6 inspections on file for Synthes, Inc. in West Chester, Pennsylvania (FEI 3005180112), from 2009 to 2013. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.

The most recent OAI inspection ended on Sep 15, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.100(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2009–2013
1 OAI · 2 VAI · 3 NAI
Inspections with 483 observations shown
3 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been documented.

  2. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation Codes

      An individual medical device manufacturer report submitted per FDA Form 3500A did not include in Block H the evaluation codes (including event codes, method of evaluation, result and conclusion codes).

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Reporting Information Reasonably Known

      An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.

    4. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 5 more citations
    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

    3. Records of complaint investigation

      Records of complaint investigations do not include required information.

    4. Records of MDR Investigation

      Investigation records of MDR reportable complaints do not include required information.

    5. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

  5. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.

  • 21 CFR 820.100Corrective and preventive action3 of 3
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
  • 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 3
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 3
  • 21 CFR 820.198Complaint files1 of 3

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005180112 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005180112. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.