Skip to content

Biochem Laboratory, Inc.

56 Park Ave, Lyndhurst, New Jersey, United States

FEI 3005094136DUNS 832494913

FDA has 6 inspections on file for Biochem Laboratory, Inc. in Lyndhurst, New Jersey (FEI 3005094136), from 2010 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 211, the manufacturing rules for finished drugs. It is listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jun 15, 2011. FDA issued a warning letter on (CDER). No court action or import refusal is filed under this FEI.

Eight regulations were cited at more than one inspection; they are listed below.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2027. Its registrant is Biochem Laboratory, Inc (D-U-N-S 832494913).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2010–2025
1 OAI · 5 VAI · 0 NAI
Inspections with 483 observations shown
6 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
4 requests, latest May 5, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling and drug products conform to appropriate standards of identity, strength, quality and purity.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Instruments, apparatus, et. al. not meeting specs

      The use of instruments not meeting established specifications was observed.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Valid stability test methods

      The written stability program for drug products does not include specific test methods.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Test methods

      The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented.

    2. Calibration - at intervals, written program, remedial action

      The calibration of instruments is not done at suitable intervals in accordance with an established written program and with provisions for remedial action in the event accuracy and/or precision limits are not met.

    3. Testing and standardization of standards et. al.

      Laboratory records do not include complete records of any testing and standardization of laboratory reference standards, reagents, and standard solutions.

    4. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.

    5. Calculations performed are in the records

      Laboratory records do not include a record of all calculations performed in connection with the test.

    Show 1 more citation
    1. Following/documenting laboratory controls

      Established standards, test procedures, and laboratory control mechanisms are not followed and documented at the time of performance.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Testing and standardization of standards et. al.

      Laboratory records do not include complete records of any testing and standardization of laboratory reference standards, reagents, and standard solutions.

    2. Valid stability test methods

      The written stability program for drug products does not include reliable, meaningful, and specific test methods.

    3. Defined areas of adequate size for operations

      The control systems necessary to prevent contamination or mix-ups are deficient.

    4. Training--operations, GMPs, written procedures

      Employees are not given training in the particular operations they perform as part of their function and current good manufacturing practices.

    5. Calibration - at intervals, written program, remedial action

      The calibration of instruments is not done at suitable intervals in accordance with an established written program and with provisions for remedial action in the event accuracy and/or precision limits are not met.

    Show 7 more citations
    1. Signatures and dates--person who performs test

      Laboratory records do not include the initials or signature of the person who performs each test.

    2. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.

    3. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate standards and test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.

    4. Following/documenting laboratory controls

      Established standards, test procedures, and laboratory control mechanisms are not followed and documented at the time of performance.

    5. Calculations performed are in the records

      Laboratory records do not include a record of all calculations performed in connection with the test.

    6. Computer control of master formula records

      Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.

    7. Test methods

      The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. GMP Training Frequency

      GMP training is not conducted on a continuing basis and with sufficient frequency to assure that employees remain familiar with CGMP requirements applicable to them.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on May 5, 2026. Each one asked for a Form 483.

  • Inspection ending Jul 24, 2025VAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 10, 2025. FDAgent holds no copy that FDA posted.

    1. May 5, 2026FDA-FOIA-2026-4431Form 483· 1 of 5 sitesRecords released
    2. Oct 7, 2025FDA-FOIA-2025-11653Form 483· 1 of 5 sitesRecords released
    3. Sep 30, 2025FDA-FOIA-2025-11425Form 483Records released
  • Inspection ending Mar 26, 2019VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 23, 2023FDA-FOIA-2023-9318Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDER · DrugsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

FDA drug establishment registration

FDA registration search
Registered operations
Analysis
Registration runs through
Dec 31, 2027
Registrant
Biochem Laboratory, IncD-U-N-S 832494913

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005094136 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005094136. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.