MedMira Laboratories Inc.
Suite 1 Unit 1, 155 Chain Lake Dr, Halifax, Canada
FEI 3003595936
FDA has 3 inspections on file for MedMira Laboratories Inc. in Halifax, Canada (FEI 3003595936), from 2010 to 2015. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
9 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2010–2015
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 9 refusals
- Import alerts
- 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Device identification, control numbers
The device history record does not include certain device identifications and control numbers used.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Equipment Installation, Placement, Specified Requirements
Appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Show 5 more citations
Device identification and control number
Records of complaint investigations do not include device identifications and control numbers used.
Uniform and timely processing
Complaint handling procedures have not been defined and completed to ensure that all complaints are processed in a uniform and timely manner.
Appropriateness
There is no information to support the appropriateness of the statistical techniques used.
Lack of or inadequate DHR procedures
Procedures have not been completed and implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Lack of or inadequate procedures for handling
Procedures have not been implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.140Handling1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 820.250Statistical techniques1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3003595936, one row per refused product line.
9 refusals from 2026-04-30 to 2026-09-03 · charged NOT LISTED on 9, FRNMFGREG on 9 · no FDA sample analysis
FDA firm profile- 2026-09-03
RESEARCH USE ONLY/PATHOLOGY88OTU
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports1FX-5837321-7/10002/1
- 2026-06-22
ANTIBODY TO HTLV-1, ELISA81MTP
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Northern Border Imports1FX-1744231-0/10002/1
- 2026-06-22
ANTIBODY TO HTLV-1, ELISA81MTP
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Northern Border Imports1FX-1737399-4/10002/1
- 2026-06-05
TEST, HIV DETECTION83MZF
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports1FX-1744711-1/10002/1
- 2026-06-05
TEST, HIV DETECTION83MZF
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports1FX-1744485-2/10002/1
Show 4 more
- 2026-06-02
TEST, HIV DETECTION83MZF
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports1FX-1117944-7/10002/1
- 2026-05-06
TEST, HIV DETECTION83MZF
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports1FX-7410182-4/10002/1
- 2026-04-30
TEST, HIV DETECTION83MZF
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports1FX-6425892-3/10002/1
- 2026-04-30
TEST, HIV DETECTION83MZF
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports1FX-5807694-3/10002/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2021-01-29
- FDA’s entry
- MedMira Laboratories Inc. · Suite 1 Unit 1, 155 Chain Lake Dr, Halifax, Nova Scotia CANADA
- FDA’s inspection record for FEI 3003595936
- MedMira Laboratories Inc. · Suite 1 Unit 1, 155 Chain Lake Dr, Halifax, Canada
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
1 product
- 83 Q - - KOReagent, coronavirus serological
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003595936 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003595936. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.