Neusoft Medical Systems Co., Ltd.

No. 177-1 Chuangxin Road, Hunnan New District, Shenyang, China

FEI 3003575350

FDA has 5 inspections on file for Neusoft Medical Systems Co., Ltd. in Shenyang, China (FEI 3003575350), from 2010 to 2019. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 1020.30(k), 21 CFR 820.90(a) and 21 CFR 1002.20(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2010–2019
0 OAI · 2 VAI · 3 NAI
Inspections with 483 observations shown
3 of 5
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Compliance measurements--diagnostic source assembly

      Compliance for radiation leakage from the diagnostic source assembly was not determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Failure to report

      You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.

    2. Compliance measurements--diagnostic source assembly

      Compliance for radiation leakage from the diagnostic source assembly was not determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.

    3. Incorrect translation to production specifications

      The device design was not correctly translated into production specifications.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    Show 2 more citations
    1. Justification for not reporting

      A justification for not reporting the correction or removal action to FDA that included conclusions and reviews by a designated person was not included in the record.

    2. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

  4. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Dealers, distributors, purchasers not notified

      You did not notify the distributors, purchasers, and subsequent transferees of an electronic product which was defective .

    2. CAP not submitted

      In a situation where an electronic product you have manufactured has a defect subject to the notification requirements of 21 CFR 1003.10(b), you have not submitted a plan to the Secretary to remedy the defect without charge (including reimbursement of any transportation expenses), replace the product with a like or equivalent one which, or refund the cost of the product to the purchaser.

    3. Failure to report

      You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.

    4. FDA not notified of defect or noncompliance

      You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced and distributed.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.

Show all 10
  • 21 CFR 1003.10Discovery of defect or failure of compliance by manufacturer; notice requirements1 of 3
  • 21 CFR 1004.1Manufacturer's obligation to repair, replace, or refund cost of electronic products1 of 3

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003575350 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003575350. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.