CBR Systems Inc
6550 S Bay Colony Dr Ste 160, Tucson, Arizona, United States
FEI 3003374183
FDA has 4 inspections on file for CBR Systems Inc in Tucson, Arizona (FEI 3003374183), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection, all under 21 CFR Part 1271, the rules for human cell and tissue products. They are listed below. One of the five FOIA requests for this site has asked FDA for this inspection's records.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 1271.230(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2024
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 5 requests, latest Sep 11, 2024
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Use prior to verification
Supplies and reagents were used before they were verified to meet specifications designed to prevent the introduction, transmission, or spread of communicable disease.
Adequate control, proper conditions
Environmental conditions existed in which contamination or cross contamination of HCT/Ps or equipment could occur, and environmental conditions were not adequately controlled and proper conditions for operations were not provided.
Verification of label accuracy, legibility, integrity
Labeling procedures did not include verification of label accuracy.
Documentation/approval prior to implementation
Computer software validation and verification activities and results have not been approved prior to implementation.
Shipping conditions appropriate
Appropriate shipping conditions were not maintained and followed for each type of HCT/P.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Actions verified, short and long term solutions
Corrective actions relating to core CGTP requirements have not been verified to ensure effectiveness and compliance with CGTP.
Process validation procedures
Procedures to validate and approve processes that cannot be fully verified by inspection and tests were not defined.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Monitoring-microorganisms where appropriate
Environmental conditions were not monitored for microorganisms.
Validation & approval--established procedures
Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.
Risk factors, leukocyte-rich cells or tissues
Donors of leukocyte-rich cells or tissues were not screened by a review of relevant medical records for clinical evidence of cell-associated communicable disease agents and diseases.
Causing contamination, increasing risks
HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on Sep 11, 2024.
Inspection ending Jan 17, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- May 17, 2024FDA-FOIA-2024-4468EIRClosed
Inspection date not given
4 requests. None of FDA’s log entries for these requests say records were released.
- Sep 11, 2024FDA-FOIA-2024-8114EIRClosed
- Jul 17, 2024FDA-FOIA-2024-6269Form 483Closed
- May 13, 2024FDA-FOIA-2024-4300Form 483Closed
- Mar 26, 2024FDA-FOIA-2024-2900Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 1271.160Establishment and maintenance of a quality program2 of 3
- 21 CFR 1271.195Environmental control and monitoring2 of 3
- 21 CFR 1271.230Process validation2 of 3
- 21 CFR 1271.75How do I screen a donor?1 of 3
- 21 CFR 1271.210Supplies and reagents1 of 3
- 21 CFR 1271.220Processing and process controls1 of 3
- 21 CFR 1271.250Labeling controls1 of 3
- 21 CFR 1271.265Receipt, predistribution shipment, and distribution of an HCT/P1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003374183 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003374183. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.