University Of Colorado Hospital Clinical Laboratory

12425 E 16th Ave Aip 3 Room 2880, Aurora, Colorado, United States

FEI 3003233482

FDA has 7 inspections on file for University Of Colorado Hospital Clinical Laboratory in Aurora, Colorado (FEI 3003233482), from 2011 to 2024. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA issued a warning letter on (CBER). No court action or import refusal is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
7, 2011–2024
0 OAI · 2 VAI · 5 NAI
Inspections with 483 observations shown
2 of 7
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  2. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  3. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  4. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  5. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the processing, storage, and distribution of blood and blood components for homologous transfusion are not always maintained and followed.

    2. Controlling storage temperatures

      The standard operating procedure fails to include a written description of the storage temperatures and methods of controlling storage temperatures for all blood products and reagents

    3. Required records

      Failure to maintain storage and distribution records.

    4. Person performing, test results, interpretation

      Records fail to be as detailed as necessary so as to provide a complete history of the work performed.

    5. Thorough investigations

      Failure to make a record of the conclusions and follow-up of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.

  7. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products21 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the processing, storage, and distribution of each unit of blood and blood components so that all steps can be clearly traced.

    2. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.

    3. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion and autologous transfusion are not always maintained and followed.

    4. Quality control

      Failure to test each month (of manufacture) four units prepared from different donors at the end of the storage period for platelet count, pH of not less than 6.0 measured at the storage temperature of the unit, and actual plasma volume.

    5. Donations in less than eight weeks

      A person served as a source of blood more than once in 8 weeks and was not examined at the time of donation and certified by a physician to be in good health as indicated in part in 21 CFR 640.3(b).

    Show 16 more citations
    1. Determination of lot numbers and supplies

      Appropriate records are not available to determine the lot numbers of supplies used for specific units of the final product.

    2. Person performing, test results, interpretation

      Records fail to identify the person performing the work and include dates of the various entries so as to provide a complete history of the work performed.

    3. Required records

      Failure to maintain donor, storage and distribution, and general records.

    4. Following manufacturer's instructions

      Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.

    5. Donor deferral

      The standard operating procedure fails to include a written description of the donor deferral process.

    6. Qualifications of donor - disease transmissible by blood

      Failure to determine whether each donor is free from any disease transmissible by blood as determined by history and examinations.

    7. Record review prior to release

      All records pertinent to a lot or unit were not reviewed before the release or distribution of a lot or unit of final product.

    8. Autologous donation - label

      Failure to label autologous donations with the appropriate caution statement.

    9. Donor criteria

      The standard operating procedure fails to include written descriptions of criteria used to determine donor suitability, including acceptable medical history criteria.

    10. History of viral hepatitis - close contact

      An individual was used as a source of Blood who had a history of close contact within twelve months of donation with an individual having viral hepatitis.

    11. input/output verification

      Input to and output from the computer, related systems of formulas, and records or data are not checked for accuracy.

    12. Eligibility of platelet donors

      All plateletpheresis donors did not meet the criteria for suitability.

    13. Qualifications of responsible personnel

      The personnel responsible for the collection and processing of blood or blood components are not adequate in training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.

    14. Equipment calibration frequency

      Equipment used in the processing of blood and blood components is not observed, standardized and calibrated with at least the frequency required.

    15. Testing - inspection

      Failure to visually inspect blood during storage and immediately prior to issue for abnormal color, physical appearance, or indication or suspicion of microbial contamination.

    16. Calibration/Inspection/Checking not done

      Routine calibration and inspection of mechanical equipment is not performed according to a written program designed to assure proper performance.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 7 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCBER · BiologicsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003233482 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003233482. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.