Skip to content

Radiology Information Systems, Inc.

43676 Trade Center Pl Ste 100, Dulles, Virginia, United States

FEI 3002952981

FDA has 4 inspections on file for Radiology Information Systems, Inc. in Dulles, Virginia (FEI 3002952981), from 2015 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2015–2025
0 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Aug 13, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    2. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Design validation - software validation not performed

      Validation of device software is inadequate and is incomplete.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    5. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    Show 2 more citations
    1. DHR - UDI not included

      The device history record does not include, or refer to the location of any unique device identifier (UDI) or universal product code (UPC).

    2. Procedures not adequately established or maintained

      Document control procedures have not been adequately established and maintained.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    2. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been established.

    3. Design validation - software validation not performed

      Validation of device software is inadequate and is incomplete .

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Aug 13, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 18, 2025VAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 12, 2025. FDAgent holds no copy that FDA posted.

    1. Aug 5, 2025FDA-FOIA-2025-9490Form 483· 1 of 5 sitesRecords released
  • Inspection ending Jul 15, 2015VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 13, 2026FDA-FOIA-2026-7823Form 483No records found

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002952981 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002952981. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.