Puretone Ltd.
Unit 9-10, Henley Business Park; Trident Close, Rochester, United Kingdom
FEI 3002808722
FDA has one inspection on file for Puretone Ltd. in Rochester, United Kingdom (FEI 3002808722). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 10 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
37 import refusals are filed under this FEI, the latest on . FDA’s Red Lists for Import Alerts 89-04 and 89-08 name this firm at this address. FDA printed no FEI on the entries. No court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2008
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- 37 refusals
- Import alerts
- 89-04, 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Training records
Employee training is not fully documented.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Mixups
Labeling and packaging operations were not controlled adequately to prevent labeling mix-ups.
Retention period
Required records are not retained for the design and expected life of the device.
Show 5 more citations
Design review procedures - personnel
Procedures were not established for ensuring that participants at each design review include an individual who does not have direct responsibility for the design stage being reviewed.
Lack of or inadequate procedures for labeling
Labeling procedures were not defined.
Investigation procedures
The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not complete.
Documentation
Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, and implementation of corrective and preventive actions.
Training
Employees have not been adequately trained.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.120Device labeling1 of 1
- 21 CFR 820.180General requirements1 of 1
- 21 CFR 820.184Device history record1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002808722, one row per refused product line.
37 refusals from 2017-01-06 to 2020-03-12 · charged DEVICEGMPS on 32, FRNMFGREG on 12 · no FDA sample analysis
FDA firm profile- 2020-03-12
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-6021906-9/11/1
- 2020-03-12
PROTECTOR, HEARING (INSERT)77EWD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-5740915-2/11/1
- 2018-08-31
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6631544-9/21/1
- 2018-08-24
HEARING-AID, AIR-CONDUCTION77ESD
DE IMP GMP 801(a)(1); NON CONFORMING MANUFACTURING PRACTICES
Division of Southeast Imports799-6142725-6/21/1
- 2018-08-22
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6301321-1/21/1
Show 32 more
- 2018-08-02
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6560179-9/21/1
- 2018-07-20
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6631935-9/21/1
- 2018-07-03
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6303256-7/21/1
- 2018-07-03
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6303161-9/21/1
- 2018-06-22
HEARING-AID, AIR-CONDUCTION77ESD
DEVGMPS 520(f); 801(a)(1)
Division of Southeast Imports799-6331096-3/21/1
- 2018-05-29
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-5254300-4/41/1
- 2018-05-29
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5227252-1/31/2
- 2018-05-10
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-5227242-2/31/2
- 2018-05-10
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-5227242-2/31/1
- 2018-04-18
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5341009-6/21/2
- 2018-04-05
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-4808794-2/31/1
- 2018-03-29
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-4589682-4/31/3
- 2018-03-20
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5338146-1/41/1
- 2018-03-20
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5338146-1/31/1
- 2018-03-20
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5227352-9/41/1
- 2018-03-20
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5227352-9/31/1
- 2017-08-23
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1596778-7/11/1
- 2017-08-23
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1595591-5/11/1
- 2017-08-23
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1523288-5/31/2
- 2017-08-23
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1429818-4/21/1
- 2017-08-09
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1490042-5/21/3
- 2017-08-09
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1381688-7/51/1
- 2017-08-09
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-1303132-1/11/1
- 2017-07-31
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-1248995-9/11/1
- 2017-07-31
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-1232956-9/11/1
- 2017-07-19
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-1174285-3/41/2
- 2017-06-19
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-1236928-4/31/2
- 2017-03-31
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
New Orleans District Office799-9724359-2/11/1
- 2017-02-15
HEARING-AID, AIR-CONDUCTION77ESD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
New Orleans District Office799-8935473-8/11/1
- 2017-01-06
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Detroit District Office799-8459998-0/11/3
- 2017-01-06
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Detroit District Office799-8459998-0/11/2
- 2017-01-06
HEARING-AID, AIR-CONDUCTION77ESD
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Detroit District Office799-8459998-0/11/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2009-09-16
- FDA’s entry
- Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, Kent UNITED KINGDOM
- FDA’s inspection record for FEI 3002808722
- Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, United Kingdom
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
2 products
- 77 E - - SDHearing-Aid, Air-Conduction
- 77 K - - LWMasker, Tinnitis
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2009-09-16
- FDA’s entry
- Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, Kent UNITED KINGDOM
- FDA’s inspection record for FEI 3002808722
- Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, United Kingdom
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
1 product
- 77 K - - LWMasker, Tinnitis
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002808722 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002808722. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.