Skip to content

Puretone Ltd.

Unit 9-10, Henley Business Park; Trident Close, Rochester, United Kingdom

FEI 3002808722

FDA has one inspection on file for Puretone Ltd. in Rochester, United Kingdom (FEI 3002808722). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 10 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

37 import refusals are filed under this FEI, the latest on . FDA’s Red Lists for Import Alerts 89-04 and 89-08 name this firm at this address. FDA printed no FEI on the entries. No court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2008
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
37 refusals
Import alerts
89-04, 89-08Red List
Matched on FDA’s firm name and street address

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ID label, labeling

      The device history record does not include the primary identification label and labeling for each device.

    2. Training records

      Employee training is not fully documented.

    3. Acceptance records

      The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.

    4. Mixups

      Labeling and packaging operations were not controlled adequately to prevent labeling mix-ups.

    5. Retention period

      Required records are not retained for the design and expected life of the device.

    Show 5 more citations
    1. Design review procedures - personnel

      Procedures were not established for ensuring that participants at each design review include an individual who does not have direct responsibility for the design stage being reviewed.

    2. Lack of or inadequate procedures for labeling

      Labeling procedures were not defined.

    3. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not complete.

    4. Documentation

      Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, and implementation of corrective and preventive actions.

    5. Training

      Employees have not been adequately trained.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002808722, one row per refused product line.

37 refusals from 2017-01-06 to 2020-03-12 · charged DEVICEGMPS on 32, FRNMFGREG on 12 · no FDA sample analysis

FDA firm profile
  • 2020-03-12

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6021906-9/11/1

  • 2020-03-12

    PROTECTOR, HEARING (INSERT)77EWD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5740915-2/11/1

  • 2018-08-31

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6631544-9/21/1

  • 2018-08-24

    HEARING-AID, AIR-CONDUCTION77ESD

    DE IMP GMP 801(a)(1); NON CONFORMING MANUFACTURING PRACTICES

    Division of Southeast Imports799-6142725-6/21/1

  • 2018-08-22

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6301321-1/21/1

Show 32 more
  • 2018-08-02

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6560179-9/21/1

  • 2018-07-20

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6631935-9/21/1

  • 2018-07-03

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6303256-7/21/1

  • 2018-07-03

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6303161-9/21/1

  • 2018-06-22

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVGMPS 520(f); 801(a)(1)

    Division of Southeast Imports799-6331096-3/21/1

  • 2018-05-29

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5254300-4/41/1

  • 2018-05-29

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-5227252-1/31/2

  • 2018-05-10

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5227242-2/31/2

  • 2018-05-10

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5227242-2/31/1

  • 2018-04-18

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-5341009-6/21/2

  • 2018-04-05

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-4808794-2/31/1

  • 2018-03-29

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-4589682-4/31/3

  • 2018-03-20

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-5338146-1/41/1

  • 2018-03-20

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-5338146-1/31/1

  • 2018-03-20

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-5227352-9/41/1

  • 2018-03-20

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-5227352-9/31/1

  • 2017-08-23

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1596778-7/11/1

  • 2017-08-23

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1595591-5/11/1

  • 2017-08-23

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1523288-5/31/2

  • 2017-08-23

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1429818-4/21/1

  • 2017-08-09

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1490042-5/21/3

  • 2017-08-09

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1381688-7/51/1

  • 2017-08-09

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1303132-1/11/1

  • 2017-07-31

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-1248995-9/11/1

  • 2017-07-31

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-1232956-9/11/1

  • 2017-07-19

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-1174285-3/41/2

  • 2017-06-19

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-1236928-4/31/2

  • 2017-03-31

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    New Orleans District Office799-9724359-2/11/1

  • 2017-02-15

    HEARING-AID, AIR-CONDUCTION77ESD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    New Orleans District Office799-8935473-8/11/1

  • 2017-01-06

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Detroit District Office799-8459998-0/11/3

  • 2017-01-06

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Detroit District Office799-8459998-0/11/2

  • 2017-01-06

    HEARING-AID, AIR-CONDUCTION77ESD

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Detroit District Office799-8459998-0/11/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2009-09-16

    FDA’s entry
    Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, Kent UNITED KINGDOM
    FDA’s inspection record for FEI 3002808722
    Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, United Kingdom

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    2 products

    • 77 E - - SDHearing-Aid, Air-Conduction
    • 77 K - - LWMasker, Tinnitis
  • 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov

    Date published 2009-09-16

    FDA’s entry
    Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, Kent UNITED KINGDOM
    FDA’s inspection record for FEI 3002808722
    Puretone Ltd. · Unit 9-10, Henley Business Park; Trident Close, Rochester, United Kingdom

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    1 product

    • 77 K - - LWMasker, Tinnitis

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002808722 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002808722. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.