FDA has one inspection on file for Moeller Medical GmbH in Fulda, Germany (FEI 3002807634). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request asked FDA for the 483 from a later inspection, ending Mar 30, 2023, that FDA's Data Dashboard doesn't list yet.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2014
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Feb 17, 2026
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - validation vs. verification
Procedures were not established to address when verification of design changes is sufficient in lieu of validation prior to their implementation.
Management review - Lack of or inadequate procedures
Procedures for management review are not complete.
Design input procedures - addressing conflicting requirement
Procedures for addressing incomplete, ambiguous, or conflicting design input requirements were not established and complete.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures were not established and complete for monitoring and control of process parameters for validated processes.
Verification vs validation
There is no corrective and preventive action procedure to determine when verification can be conducted in lieu of validation.
Show 3 more citations
Design validation - production units
The design was not validated using production units or their equivalents.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Supplier notification of changes
There is no clear agreement from the suppliers, contractors, and consultants that they will notify you of changes in the product or service.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Feb 17, 2026. It asked for a Form 483.
Inspection ending Mar 30, 2023
1 request. Not listed in FDA’s Data Dashboard yet.
- Feb 17, 2026FDA-FOIA-2026-1644Form 483· 1 of 2 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.184Device history record1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002807634, one row per refused product line.
1 refusal on 2017-05-26 · charged NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2017-05-26
ELECTRODE, NEEDLE, DIAGNOSTIC ELECTROMYOGRAPH89IKT
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports261-0367051-4/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002807634 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002807634. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.