Skip to content

Oncology Data Systems, Inc.

1601 SW 89th St Ste E100, Oklahoma City, Oklahoma, United States

FEI 3001451801

FDA has 3 inspections on file for Oncology Data Systems, Inc. in Oklahoma City, Oklahoma (FEI 3001451801), from 2008 to 2015. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2008–2015
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Change records, content

      Records of changes did not include a description of the change and identification of the affected documents .

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of procedures, or not maintained

      Document control procedures were not implemented.

    2. Design input approval documentation

      The approval of design input requirements (including the date and signature of the individual approving the requirements) was not documented.

    3. No release signature

      Finished devices were released for distribution without signature of the individual designated to authorize such release.

    4. Establish the design and development plan

      A design and development plan has not been established.

    5. Design output - review and approval

      The approval of design output (including the date and signature of the individual approving the output) was not documented.

    Show 2 more citations
    1. Lack of or inadequate DHR procedures

      Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    2. Examination for accuracy

      Labeling was not sufficiently examined by a designated individual for accuracy including control numbers before release.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001451801 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001451801. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.