FDA has 13 inspections on file for Otsuka Pharmaceutical Development & Commercialization, Inc in Rockville, Maryland (FEI 3000718406), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.22, 21 CFR 820.30(g) and 21 CFR 820.198(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 13, 2009–2025
- 0 OAI · 5 VAI · 8 NAI
- Inspections with 483 observations shown
- 4 of 13
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
13 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Design validation - software validation not performed
Validation of device software is inadequate.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Investigation of device failures
Complaints involving the possible failure of a device and labeling to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Bioresearch MonitoringCitations not published
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sponsor device disposition records inadequate
Records of disposition of a device which describe the batch number or code marks of any devices returned to the sponsor and disposed of by the investigator or another person and the reasons for and method of disposition reports are not all accurate and complete.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.
Processing MDRs {see also 803, 804}
Complaint handling procedures have not been established, documented, and implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.
Management review - Lack of or inadequate procedures
Procedures for management review are not implemented.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Complete files maintained
Complete complaint files are not maintained.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 13 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 812.140Records1 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.100Corrective and preventive action1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000718406 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000718406. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.