Siemens Medical Solutions USA, Inc.
685 E Middlefield Rd, Mountain View, California, United States
FEI 2936884
FDA has 4 inspections on file for Siemens Medical Solutions USA, Inc. in Mountain View, California (FEI 2936884), from 2011 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection: 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below. One of the four FOIA requests for this site asked FDA for the records from a later inspection, ending Apr 6, 2017, that FDA's Data Dashboard doesn't list yet.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.198(b).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2011–2016
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Nov 18, 2021
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
Show 1 more citation
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Nov 18, 2021.
Inspection ending Apr 6, 2017
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Nov 1, 2021FDA-FOIA-2021-7444Inspection recordsClosed over fees
Inspection ending May 15, 2015
2 requests. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Nov 18, 2021FDA-FOIA-2021-7945Inspection recordsMerged with another request
- Nov 18, 2021FDA-FOIA-2021-7944Inspection recordsClosed over fees
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Jan 7, 2021FDA-FOIA-2021-172Form 483No records found
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files3 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 3
- 21 CFR 806.10Reports of corrections and removals1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
Other FEIs recorded as Siemens Medical Solutions USA, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Siemens Medical Solutions USA, Inc.FEI 3023245Issaquah, Washington, United States
Last inspected 2024-01-31NAI
Siemens Medical Solutions USA, Inc.FEI 1000514697Hoffman Estates, Illinois, United States
Last inspected 2019-06-10NAI
Siemens Medical Solutions USA, Inc.FEI 1034973Knoxville, Tennessee, United States
Last inspected 2019-06-07NAI
Siemens Medical Solutions USA, Inc.FEI 3006365753Rockford, Tennessee, United States
Last inspected 2019-06-06NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2936884 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2936884. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.