Catalent Pharma Solutions, LLC
2725 Scherer Dr N, Saint Petersburg, Florida, United States
FEI 1811396DUNS 119120716
Company: Catalent Pharma Solutions Inc. · Linked by exact D-U-N-S match (796436181) in FDA registration data
FDA has 10 inspections on file for Catalent Pharma Solutions, LLC in Saint Petersburg, Florida (FEI 1811396), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued. Four of the eight FOIA requests for this site have asked FDA for this inspection's 483.
4 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling and analysis, through Dec 31, 2027. Its registrant, Catalent Pharma Solutions Inc. (D-U-N-S 796436181), has registered 21 other sites.
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 10, 2009–2026
- 0 OAI · 8 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 10
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 8 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- 4 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written program not followed
The written stability testing program is not followed.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for non-sterile drug products
Procedures designed to prevent objectionable microorganisms in drug products not required to be sterile are not established and followed.
Quality control unit review of records
Drug product production and control records, are not reviewed by the quality control unit to determine compliance with all established, approved written procedures before a batch is released or distributed.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
Written in-process control procedures
Written procedures are not followed that describe the examinations to be conducted on appropriate samples of in-process materials of each batch.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance14 citations published
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate investigations and documentation
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, A. Investigation PR 440585 opened on 2/11/14 to investigate the presence of foreign dark matter in a gel receiver was not adequate for the following reasons: a) The foreign dark matter was not identified but was described as dark particles coming down from the (b)(4) shaft. The work order showed the replacement of the shaft seal of the (b)(4); however, the investigation did not extend to other batches (b)(4) with the same (b)(4) prior to detecting the leak on 1/30/14. b) The investigation did not extend to the previous seal change for this (b)(4) which had occurred two (2) days earlier on 1/28/14. An interview with the mechanic during the FDA inspection revealed that the (b)(4) had leaked all of its lubricant (b)(4) due to a broken seal; the seal was replaced on 1/28/14 but a manufacturing investigation was not opened to assess the impact of this lubricant on the quality of the gel batches previously (b)(4) with this (b)(4). c) A total of 1 (b)(4) sub-batches of gel mass were (b)(4) with this (b)(4) between 1/28/14 and 1/30/14 when it was observed leaking again. The impact to the quality of these batches was not assessed. d) SOP OTS-QA-0038 "Investigations" was not followed in that the investigation was opened eleven (11) days later and not within (b)(4) of the incident. B. The investigations for Valproic Acid Capsules, USP 250 mg (detailed in PR#s 408449, 414181, 478203, 488400, 500065, 513500) are deficient in that a detailed root cause analysis has not been conducted to determine the potential root cause of the multiple complaints received for blue spots on the finished capsules. Since 2013, a total of six (6) complaints were received by your firm for released commercial bulk lots 1291888, 1391388, 1391390, and 1460931. Our review of the associated investigations revealed that you have not documented detailed assessment of potential root causes that include, but not limited to; (a) potential impact of chipped and worn-out areas with blue paint on the encapsulation machines that are in close proximity to the product-contact surface areas, b) potential impact of the use of blue nitrile gloves and other gloves that are routinely used by the operators during routine manufacturing, etc. In addition, finished product retains representing the (b)(4) of the packaging runs were evaluated as part of the investigation for lot #1291888. Evaluation of the retain samples representing the (b)(4) of the packaging runs (for the same lot) were not conducted before they were discarded (approximately 1 year after the finished product expiry date). In the absence of a documented and detailed root cause analysis, your common conclusion on all the investigations that the issue is not related to manufacturing operations at your firm is unsubstantiated and potentially incomplete. C. PR# 441713 was generated on 2/9/2014, when temperature of (b)(4) refrigerator, (b)(4) was found to be out of acceptable operating range. Your firm issued a work order #1457847 to investigate and repair the refrigerator. Review of temperature data recorder indicated that refrigerator (b)(4) was repaired and started working on 3/25/2014. As part of this investigation, your firm failed to: a. Identify what material or products were stored in the refrigerator when incident happened. b. Identify the equipment (refrigerator) where impacted materials/product were moved from affected refrigerator. c. Assess any damage to the materials/products stored in affected refrigerator.
- Observation 2
Field Alert Report submission failures
An NDA-Field Alert Report was not submitted within (b)(4) of receipt of information concerning significant chemical, physical, or other change or deterioration in a distributed drug product. Specifically, since 2013 your firm has received six (6) product complaints associated with foreign material for marketed batches of Valproic Acid Capsules, USP 250 mg (product number 10258780). a) An NDA-Field Alert was submitted for ONLY two (2) out of the six (6) complaints received as detailed below: Date Received Lot Number Complaint Field Alert Status 6/20/2013 1291888 Blue Spots on Capsules Initial Field Alert 7/31/2013 1291888 Blue Spots on Capsules (2nd 08/05/2013 complaint for the same lot) 10/31/2014 1391388 Blue Spots on Capsules Field Alert not submitted 01/16/2015 1391390 Blue Spots on Capsules Field Alert not submitted 04/06/2015 Unknown Capsules with Speckling Blue dye Field Alert not submitted 08/21/2015 1460931 Blue Spots on Capsules Field Alert not submitted b) The initial Field Alert for the complaint associated with lot number 1291888 was not submitted within (b)(4) of receipt of information. Specifically, the 1st complaint was received on 06/20/2013, and the initial Field Alert was not submitted until after the 2nd complaint was received for the same lot on 7/31/2013. c) You have not established written procedures to identify and describe the roles and responsibilities for submitting Field Alerts to the FDA.
Show observations 3 to 14
- Observation 3
Insufficient computer system controls
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. (b)(4) Specifically, your firm failed to timely validate the (b)(4). In addition, there is no traceability of these incidents as the audit trail of the (b)(4) did not record in-process check results which were automatically captured but not acknowledged by the operator. For example, during the end of the encapsulation run of Targretin softgel capsules, lot 1583104, the fill and seal thickness in-process checks taken on 3/21/16 at 20:14 hrs. were inadvertently lost due to errors by the operator and his supervisor. A review of the audit trail showed no evidence of in-process checks performed at that time.
- Observation 4
Insufficient production/process control procedures
There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess. Specifically, your Site Validation Master Plan (Document No. V-133A.15, Effective: 07/02/15) states (b)(4) that (b)(4) and evaluation therein serves as a method for Continuous Process Verification. The content of your current (b)(4) was found to be deficient in that it only trends and compares critical quality attributes (mainly Assay data) during two consecutive review periods. The (b)(4) does not take into account any potential process drifts that may have occurred since the original validated status of the product. For example, during the inspection we discovered a potential trend in the Assay results for Lubiprostone, 24 mcg Capsules (product number: 10258848) in approximately 99 commercially released batches since 2008. Additionally, a potentially more significant trend in the Assay results was discovered during review of the Assay Data for Lubiprostone, 8 mcg Capsules (product number: 10258838) in approximately 45 commercially released batches since 2007. The trends that we discovered during the inspection were previously unknown to the firm. Trending Assay values as part of Continuous Process Verification may potentially serve as critical tool for gleaning additional process understanding for products such as Lubiprostone. This product exhibits therapeutic effect at microgram levels of the Active Pharmaceutical Ingredient (API) in the dosage form.
- Observation 5
Inadequate cleaning and validation practices
Equipment and utensils are not cleaned and maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product. Specifically, A. Your current Cleaning Validation program is deficient in that the current Clean Equipment Hold Time (detailed in Document ID: V-175A.01, Approved 07/24/2013) is not substantiated by holding the equipment in a "clean" status for the established period of study (b)(4). For example, your encapsulation stations were not held in a "clean" status to establish the clean hold time of (b)(4). Instead, the clean hold time studies are based on review of historical environmental monitoring data. This approach does not take into account the continued cleanliness of individual equipment and equipment surfaces that are at higher risk for potential microbial proliferation during the holding period. During the environmental monitoring period between January 2010 and March 2013, there were four (4) recorded events during which the microbial levels exceeded the action levels and a definitive root cause for the excursions was not identified. Additional commercial production equipment that utilized environmental monitoring to establish the Clean Hold Time but not limited to: B. During the inspection on 3/24/16 of the (b)(4), which was tagged as "Clean" and intended for the preparation of the (b)(4) softgel capsules it was observed that white powder residue remained around the edge of the (b)(4) hole after it was (b)(4) cleaned. According to SOP OTS-GEL-0047 "Operation/Cleaning (b)(4)" version 1.0, a visual inspection is required through the sight glass to confirm the equipment is clean; however, this residue could only be observed if the (b)(4) cover is opened.
- Observation 6
Quality unit procedures not followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, a) Your quality unit failed to follow SOP OTS-GM-0029 "Production Master Records - Production Events Log" in that production events/deviations reported by email or phone to QA by the (b)(4) production shifts were not adequately tracked, documented, and investigated to determine the root cause and take appropriate corrective and preventive action. b) Your firm lacks written procedures for the segregation of in-process material when a deviation occurs or the material requires further evaluation. 3 of SOP OTS-QC-0034 for lab investigation requires opening investigation within (b)(4) when lab becomes aware of the OOS results. Investigation PR# 434491 was initiated on 12/30/2013 in response to Palonosetron HCl out of specification assay results. The Quality Control Unit became aware of OOS results on 12/19/2013 but the investigation was not initiated until 12/30/13 (11 days later). d) Extension requests for investigations that are not completed within due dates are not justified in the (b)(4) system. More specifically investigation PR# 469043 was created on 8/29/2014 and was closed after 322 days on 7/17/2015. Another investigation, PR# 455373, was created on 5/22/2014 and closed after 321 days on 4/8/2015. Current version 5.0 of SOP OTS-QA-0005, (b)(4) Task System, Effective: 10/14/2015 does not require any justification for granting extension for investigations.
- Observation 7
Production procedures not followed/documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance. Specifically, during the walk-thru inspection of the East Wing Encapsulation Area on 3/21/16 around 4:00 p.m., it was observed that the softgel encapsulation machine in Station (b)(4) was partially disassembled and was tagged with "Repair in progress" and Work Order #3012261 dated 3/21/16 which stated "pump failure, gears do not engage." An interview with production personnel revealed that (b)(4) was being encapsulated when the operator heard a loud noise coming from the equipment. The encapsulator was stopped at 9:54 a.m. and the few capsules remaining on the conveyor belt were rejected. The filled capsules continued with processing even though in-process checks for fill weights had not been performed for an hour since 8:42 a.m. The encapsulation machine was partially disassembled, the gel mass and fill materials were disconnected and held, but the incident was not documented in the Production Event Log within the electronic batch record when it occurred and QA was not notified as required by SOP OTS-GM-0029 "Production Master Records- Production Events Log" when equipment malfunction or breakdown occurs. Production supervisors had stated during the walk-thru that QA notification was not required; therefore, it is unknown how many incidents of equipment malfunction/breakdown have occurred during production that have not been notified to QA for further evaluation of product impact.
- Observation 8
Inadequate supervision of weighing operations
Component weighing operations are not adequately supervised. Specifically, adequate controls for weight verification of components were not in place between 8/15/14 and 10/20/14 when a glitch was found in the (b)(4) system that interfaces with the scales and the (b)(4) display unit which resulted in weight discrepancies upon taring. According to investigation PR465555 opened on 8/1/14 to investigate OOS results for (b)(4) lots, the root cause was identified as weight discrepancies caused by connectivity issues between the systems. As an interim corrective action, your quality unit implemented a "Production Event Advisory" on 8/15/14 to add additional steps for weight verification by a supervisor who had to initial and date every weight performed on Scale (b)(4) Room (b)(4). However, as the verification was reportedly documented on the label of each raw material which was discarded upon dispensing, your firm lacked documentation to demonstrate that every component weighed on the affected scale was verified for accuracy. The (b)(4) system was not modified and qualified until 10/20/14 to incorporate acknowledgment codes to verify scale connectivity and accuracy of weights.
- Observation 9
Deficient contamination/mix-up controls
The control systems necessary to prevent contamination or mix-ups are deficient. Specifically, on 3/24/16 a stack of unidentified trays containing yellow softgel capsules were observed inside drying tunnel in suite (b)(4). This material was not logged in the (b)(4) Clearance and Usage Log for Tunnel ID (b)(4).
- Observation 10
Inadequate employee training on SOPs
Employees are not given training in the particular operations they perform as part of their function. Specifically, upon interview of operators in the warehouse and production areas it was found that they lacked training on how to access the electronic SOPs in Documentum and therefore were not routinely referring to them for the performance of their duties. Deviations from written procedures were observed during the inspection.
- Observation 11
Deficient laboratory identification procedures
Laboratory controls do not include the establishment of scientifically sound and appropriate standards and test procedures designed to assure that components, in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, the identification test by (b)(4) performed as part of the release testing of (b)(4) active pharmaceutical ingredients (API) (i.e.) (b)(4) is not (b)(4). As part of the test, the (b)(4). The reference standard (b)(4) were not generated concomitantly and under the same conditions as the tested samples. In addition, there are no procedures for limiting the number of samples (b)(4) tested concurrently resulting in several (b)(4) printed on the same graph making it difficult to determine differences in the (b)(4) profile.
- Observation 12
Calibration procedures not followed
Procedures describing the calibration of instruments, apparatus, gauges and recording devices are deficiently written or followed. Specifically, A. Your firm did not perform all the required tests to qualify high performance liquid chromatography (HPLC) system ID (b)(4) in Sep 2015 as per the vendor requirements and specifications. More specifically, the following studies were not performed as part of qualification of (b)(4): No Test Name Set Points and Parameters/ Limits 1 Height RSD: Injection Vol. (b)(4) Column: Area RSD: (b)(4) 2 Height Carry Injection Vol. (b)(4) Column: Area Carry (b)(4) 3 Accuracy: (b)(4) On 3/23/2016, your Analytical R&D Group Leader stated that these studies were covered under other studies but there were no such comments, justification or explanation documented in equipment qualification report for this particular equipment. B. More specifically your QC and micro labs use (b)(4) micro pipettes to prepare various solutions of samples, standards, and reagents. These solutions are used to assess the quality of drug product as if it meets the specifications for commercial release. During inspection we reviewed calibration reports of following pipettes: No | Pipette Name | Serial No | Range | Cal Due ( b)(4 ) During our review, we discovered that micropipettes are shipped for calibration to a vendor in (b)(4). Vendor performs calibration and ships back these pipettes with "As Returned" data without performing any "As Received" studies. Lab management stated that micropipettes are calibrated on (b)(4) basis and lab does not perform any verification checks between the uses during the (b)(4) of use. There is no documented evidence that these pipettes were working as intended when they were originally received by vendor for calibration.
- Observation 13
Laboratory test procedure noncompliance
Established test procedures and laboratory control mechanisms are not followed and documented at the time of performance. Specifically, A. Your SOP QTS-QC-0049, Version: 12.0, Effective Date: 1/28/15, section 6.1.1 states "Each refrigerator/freezer's reading will be recorded (b)(4) and documented in the front temperature log section of the OTS-QC-0049-F01". However, the following tasks performed in QC Lab do not comply with this SOP: 1) Review of temperature data within the last four (4) months showed measurements were not taken on the following 13 days: 3/4/16, 3/10/16, 2/5/16, 1/13/16, 1/14/16, 1/12/16, 12/16/15, 12/17/15, 12/18/15, 12/20/15, 12/23/15, 12/24/15, 12/28/15. Heavy frost was observed in the freezer that is used to store reference standards. As per procedure OTS-QC-0049 (b)(4). 2) Many entries were not made on "Performed By" column on logbook of stability chamber, ID: SM1001 that is used to store products at room temperature. More specifically, on 11/24/15, and 3/2/16 stability samples for lot# (b)(4), and (b)(4) were pulled and no record was entered as to who pulled the samples. C. A review of the raw data for Palonosetron HCl in-process assay (for fill material lot# (b)(4) and FP lot# 125353) in notebook #8624, page #19, project #1459 revealed that the preparation of the (b)(4) that was used in the analysis was not documented on the notebook. Analyst performed this test on 10/29/2015 and data was reviewed on 11/02/2015. In this particular case, the analyst failed to document (b)(4) information and the reviewer failed to detect the missing information.
- Observation 14
No written adverse event procedures
Written procedures have not been developed for the receipt and evaluation of post marketing adverse drug experiences. Specifically, during the inspection, your firm provided a Quality Call Handling worksheet that outlines the handling of telephone calls involving potential Adverse Events for the products owned by Catalent. This document is deficient in that it is: (a) not a controlled document, (b) there is no review or approval information on this document to indicate that it is formally reviewed and approved by your quality unit and other groups that are responsible for collection and evaluation of potential Adverse Events, and (c) there is no documentation to indicate that the personnel involved in the receipt and evaluation of the adverse events are trained on this procedure.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Eight requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Jan 30, 2026VAI
4 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 + EIR on Mar 20, 2026. FDAgent holds no copy that FDA posted.
- May 11, 2026FDA-FOIA-2026-4805Form 483· 1 of 5 sitesWithdrawn
- Apr 6, 2026FDA-FOIA-2026-3367Form 483· 1 of 4 sitesRecords released
- Mar 20, 2026FDA-FOIA-2026-2785Form 483 + EIRRecords released
- Mar 13, 2026FDA-FOIA-2026-2566Form 483· 1 of 5 sitesRecords released
Inspection ending Jul 21, 2023
2 requests. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-6910Form 483Closed
- Oct 6, 2023FDA-FOIA-2023-8802Form 483Closed
Inspection ending Aug 28, 2019VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 6, 2023FDA-FOIA-2023-8805Form 483Closed
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Oct 2, 2024FDA-FOIA-2024-8637Form 483 + EIR· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 211.22Responsibilities of quality control unit1 of 3
- 21 CFR 211.100Written procedures; deviations1 of 3
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 3
- 21 CFR 211.113Control of microbiological contamination1 of 3
- 21 CFR 211.160General requirements1 of 3
- 21 CFR 211.166Stability testing1 of 3
- 21 CFR 211.192Production record review1 of 3
- 21 CFR 820.75Process validation1 of 3
Show all 9
- 21 CFR 820.90Nonconforming product1 of 3
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1811396, one row per refused product line.
4 refusals from 2016-10-14 to 2017-07-19 · charged UNAPPROVED on 3, NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2017-07-19
PROGESTERONE (PROGESTIN)65JCH31
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Southeast Imports231-1167502-4/211/1
- 2017-01-06
ANTI-CHOLESTEREMIC, N.E.C.61JCH99
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports231-1162416-2/171/1
- 2016-10-14
PROGESTERONE (PROGESTIN)65JCH31
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Southeast Imports231-1159559-4/9/1
- 2016-10-14
PROGESTERONE (PROGESTIN)65JCH31
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Southeast Imports231-1159559-4/10/1
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling, analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- Catalent Pharma Solutions Inc.D-U-N-S 796436181
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other sites with the same registrant
FDA’s registration file names Catalent Pharma Solutions Inc. (D-U-N-S 796436181) as the registrant for this site and 21 others. Listed: the 12 with the most warning letters, then the most recent inspections. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.
- Catalent Pharma Solutions, LLCMorrisville, North Carolina · FEI 100011091212 inspections12latestMay 22, 2026VAINone on file
- Catalent Massachusetts, LLCChelsea, Massachusetts · FEI 30116961153 inspections3latestApr 7, 2026VAINone on file
- Catalent Pharma Solutions, LLCPhiladelphia, Pennsylvania · FEI 30076170944 inspections4latestDec 17, 2025NAINone on file
- Catalent Greenville, Inc.Greenville, North Carolina · FEI 300020973513 inspections13latestOct 24, 2025VAINone on file
- Catalent Micron Technologies, Inc.Malvern, Pennsylvania · FEI 30079922667 inspections7latestJun 9, 2025VAINone on file
- Catalent Ontario LimitedStrathroy, Canada · FEI 30073254364 inspections4latestMay 2, 2025VAINone on file
- Catalent Pharma Solutions, LLCWinchester, Kentucky · FEI 100012240010 inspections10latestApr 22, 2025VAINone on file
- Catalent Maryland, Inc.Baltimore, Maryland · FEI 30155585902 inspections2latestApr 9, 2025VAINone on file
- Catalent Maryland, Inc.Harmans, Maryland · FEI 30154343014 inspections4latestNov 15, 2024NAINone on file
- CATALENT U.K. SWINDON ZYDIS LIMITEDSwindon, United Kingdom · FEI 30038125858 inspections8latestOct 10, 2024VAINone on file
- Catalent Germany Eberbach GmbHEberbach, Germany · FEI 30028080986 inspections6latestFeb 2, 2024VAINone on file
- CATALENT ARGENTINA S.A.I.C.Buenos Aires, Argentina · FEI 30042585116 inspections6latestNov 9, 2023VAINone on file
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. This site’s records are filed under FEI 1811396 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every inspection and enforcement record on this page is filed under FEI 1811396. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.