Sontec Instruments, Inc
7248 S Tucson Way, Centennial, Colorado, United States
FEI 1720747
FDA has one inspection on file for Sontec Instruments, Inc in Centennial, Colorado (FEI 1720747). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2016
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Documented evaluation
The evaluation of potential suppliers was not documented.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Design review - documentation
The design review results, including identification of the design, the date and the individual(s) performing the review , were not documented in the design history file.
Design verification - documentation
The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification , were not adequately documented in the design history file.
Show 2 more citations
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation , were not documented in the design history file.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.198Complaint files1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1720747, one row per refused product line.
1 refusal on 2024-05-23 · charged DIRECTIONS on 1, FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2024-05-23
MESH, SURGICAL, NON-ABSORBABLE, ORTHOPAEDICS, REINFORCEMENT OF TENDON87OWK
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-0832669-8/11/2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1720747 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1720747. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.