Sage Products, LLC
3909 3 Oaks Rd, Cary, Illinois, United States
FEI 1419181DUNS 054326178
FDA has 9 inspections on file for Sage Products, LLC in Cary, Illinois (FEI 1419181), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.
FDA has issued 2 warning letters under this FEI, the latest on (CDER). 3 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 211.192.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing and API manufacturing, through Dec 31, 2027. Its registrant is Sage Products LLC (D-U-N-S 054326178).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 9, 2009–2025
- 0 OAI · 5 VAI · 4 NAI
- Inspections with 483 observations shown
- 7 of 9
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest May 7, 2026
- Import refusals
- 3 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written record of investigation incomplete
Written records of investigations into unexplained discrepancies do not always include the conclusions and follow-up.
Representative Samples Criteria
The number of containers to be sampled and amount of material taken from each container is not based upon appropriate criteria.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
No written record of investigation
Written records are not always made of investigations into unexplained discrepancies.
Deviations from laboratory control requirements
Deviations from written test procedures and laboratory mechanisms are not recorded and justified.
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Establish reliability of supplier's C of A
Establishment of the reliability of the component supplier's report of analyses is deficient in that the test results are not appropriately validated at appropriate intervals.
Procedures To Be in Writing
Written procedures are lacking which describe in sufficient detail the sampling , testing , approval and rejection of components , drug product containers and closures.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Show 1 more citation
Test methods
The accuracy , sensitivity , specificity and reproducibility of test methods have not been established and documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigations of discrepancies, failures
There is a failure to thoroughly review whether or not the batch has been already distributed.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Failure to develop written procedures
Written procedures have not been developed for the surveillance and evaluation of post marketing adverse drug experiences.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on May 7, 2026. Each one asked for a Form 483.
Inspection ending Jul 11, 2025VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 9, 2025. FDAgent holds no copy that FDA posted.
- May 7, 2026FDA-FOIA-2026-4676Form 483· 1 of 4 sitesWithdrawn
- Oct 2, 2025FDA-FOIA-2025-11557Form 483· 1 of 5 sitesRecords released
- Sep 30, 2025FDA-FOIA-2025-11432Form 483Records released
Inspection ending Jun 7, 2019VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 24, 2024FDA-FOIA-2024-790Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 9 inspections, with how many of them cited each.
- 21 CFR 211.192Production record review5 of 7
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures2 of 7
- 21 CFR 211.67Equipment cleaning and maintenance1 of 7
- 21 CFR 211.80General requirements1 of 7
- 21 CFR 211.160General requirements1 of 7
- 21 CFR 211.165Testing and release for distribution1 of 7
- 21 CFR 314.80Postmarketing reporting of adverse drug experiences1 of 7
- 21 CFR 820.30Design controls1 of 7
Show all 9
- 21 CFR 820.75Process validation1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1419181, one row per refused product line.
3 refusals from 2024-02-12 to 2024-04-25 · charged NOT LISTED on 2, AGR RX on 1 · no FDA sample analysis
FDA firm profile- 2024-04-25
CETYLPYRIDINIUM CHLORIDE (ANTI-INFECTIVE, TOPICAL)62FDG07
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southwest Imports800-1109975-3/111/1
- 2024-04-25
CETYLPYRIDINIUM CHLORIDE (ANTI-INFECTIVE, TOPICAL)62FDG07
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southwest Imports800-1109975-3/101/1
- 2024-02-12
CHLORHEXIDINE HCL (ANTI-INFECTIVE, TOPICAL)62FDY08
AGR RX 801(d)(1),(2); IMPORTATION RESTRICTED
Division of Northeast Imports9SN-0007009-6/111/1
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, API manufacturing
- Registration runs through
- Dec 31, 2027
- Registrant
- Sage Products LLCD-U-N-S 054326178
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1419181 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1419181. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.