Skip to content

BSN Medical Inc

11405 N Community House Rd Ste 600, Charlotte, North Carolina, United States

FEI 1038963

FDA has 3 inspections on file for BSN Medical Inc in Charlotte, North Carolina (FEI 1038963), from 2017 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.30(g).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2017–2025
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Feb 10, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - simulated testing

      The design was not validated under actual or simulated use conditions.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation - review in response to changes or deviations

      There is no documentation of the review and evaluation of a process performed in response to changes or process deviations.

    2. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    3. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    4. Design validation - production units

      The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.

    5. Documentation

      Process validation activities and results have not been documented.

    Show 1 more citation
    1. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed and investigated where necessary.

    2. Lack of System for Determining MDR Events

      The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 10, 2026. Each one asked for a Form 483.

  • Inspection ending Sep 5, 2025VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Feb 26, 2026. FDAgent holds no copy that FDA posted.

    1. Feb 10, 2026FDA-FOIA-2026-1426Form 483· 1 of 4 sitesRecords released
    2. Nov 19, 2025FDA-FOIA-2025-13097Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1038963 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1038963. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.