Skip to content

Orfit Industries NV

Vos Veld 9a, Wijnegem, Belgium

FEI 1000227792

FDA has 4 inspections on file for Orfit Industries NV in Wijnegem, Belgium (FEI 1000227792), from 2015 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.22.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2015–2026
0 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jul 22, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.5.6

      Procedures for validation of processes lack required elements

      Procedures for validation of processes did not include defined criteria for review and approval of the processes.

    2. ISO 13485:2016 Clause 4.2.4

      Document Control Procedure Inadequate

      The document control procedure does not adequately define controls needed to review and approve documents for adequacy before use and ensure that current versions of applicable documents are available at point of use.

    3. ISO 13485:2016 Clause 7.5.1

      Production and service provision not adequate

      Production and service provision was not adequately planned, carried out, monitored, and controlled to ensure that product conforms to specification.

    4. ISO 13485:2016 Clause 7.6

      Procedures for monitoring and measuring equipment requirements not documented

      Procedures to ensure that monitoring and measurement can be carried out and are carried out in a manner that is consistent with the monitoring and measurement requirements were not adequately documented.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audits - defined intervals

      Quality audits were not performed at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.

    2. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.

    2. Quality Audits - defined intervals

      Quality audits were not performed at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.

    3. Design review - documentation

      The design review results, including the date and the individual(s) performing the review, were not documented in the design history file.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jul 22, 2026. It asked for a Form 483.

  • Inspection ending Apr 2, 2026VAI

    1 request. In the inspection history above.

    1. Jul 22, 2026FDA-FOIA-2026-6993Form 483· 1 of 2 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000227792 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000227792. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.