Skip to content

Peregrine Surgical Ltd

51 Britain Dr, New Britain, Pennsylvania, United States

FEI 1000121498

FDA has 5 inspections on file for Peregrine Surgical Ltd in New Britain, Pennsylvania (FEI 1000121498), from 2010 to 2018. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Jul 15, 2010. FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2010–2018
1 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
2 of 5
Form 483s posted by FDA
None held for this FEI
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Promotion or test marketing

      An investigational device was test marketed prior to FDA approval/clearance.

    3. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    5. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    Show 1 more citation
    1. Design review - documentation

      The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 5 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000121498, one row per refused product line.

1 refusal on 2019-11-26 · charged REGISTERED on 1 · no FDA sample analysis

FDA firm profile
  • 2019-11-26

    UNIT, PHACOFRAGMENTATION86HQC

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-6441026-6/21/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000121498 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000121498. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.