Skip to content

21 CFR 803.33: If I am a user facility, what must I include when I submit an annual report?

FDA's inspection data lists 2 Form 483 citations of 21 CFR 803.33 at 2 facilities, from July 2009 to June 23, 2014.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
2
At 2 inspections
Facilities cited
2
Distinct FEIs
Most recent citation
Jun 23, 2014
Mini Lap Technologies Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 803.33 from the Electronic Code of Federal Regulations.

§ 803.33 If I am a user facility, what must I include when I submit an annual report?

(a) You must submit to us an annual report on Form FDA 3419. You must submit an annual report by January 1, of each year. You may obtain this form on the internet at: https://www.fda.gov/media/72292/download.

(b) You must clearly identify your annual report as such. You must submit your annual report to FDA, CDRH, Medical Device Reporting, P.O. Box 3002, Rockville, MD 20847-3002. Your annual report must include:

(1) Your CMS provider number used for medical device reports, or the number assigned by us for reporting purposes in accordance with § 803.3;

(2) Reporting year;

(3) Your name and complete address;

(4) Total number of reports attached or summarized;

(5) Date of the annual report and report numbers identifying the range of medical device reports that you submitted during the report period (e.g., 1234567890-2011-0001 through 1000);

(6) Name, position title, and complete address of the individual designated as your contact person responsible for reporting to us and whether that person is a new contact for you; and

(7) Information for each reportable event that occurred during the annual reporting period including:

(i) Report number;

(ii) Name and address of the device manufacturer;

(iii) Device brand name and common name;

(iv) Product model, catalog, serial, and lot number and unique device identifier (UDI) that appears on the device label or on the device package;

(v) A brief description of the event reported to the manufacturer and/or us; and

(vi) Where the report was submitted, i.e., to the manufacturer, importer, or us.

(c) In lieu of submitting the information in paragraph (b)(7) of this section, you may submit a copy of each medical device report that you submitted to the manufacturers and/or to us during the reporting period.

(d) If you did not submit any medical device reports to manufacturers or us during the time period, you do not need to submit an annual report.

21 CFR Part 803: Medical Device Reporting, Subpart CRead on eCFR

Citations by year

Form 483 citations of 21 CFR 803.33 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 1 in 2009

Form 483 citations of 21 CFR 803.33 by inspection year
YearForm 483 citations
20091
20141

How FDA words this citation

FDA's short descriptions and citation text on these 2 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Failure to Submit Annual Reports2 citations

Citation text

  • “Annual user facility reports have not been submitted to FDA.”2 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 803.33, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Mini Lap Technologies Inc.2014-06-23 · FEI 3007123990Failure to Submit Annual Reports
  • LasikPlus2009-07-14 · FEI 3007694927Failure to Submit Annual Reports

Often cited at the same inspection

Of the 2 inspections where FDA cited 21 CFR 803.33, the share where it also cited each of these sections.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 (100%)
  • 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 (50%)
  • 21 CFR 803.30If I am a user facility, what reporting requirements apply to me?1 (50%)
  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 (50%)
  • 21 CFR 820.100Corrective and preventive action1 (50%)
  • 21 CFR 820.198Complaint files1 (50%)

Ask FDAgent about 21 CFR 803.33

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

Topics that count this section:Device CAPA and complaints

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 803.33; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.