Theken Spine LLC. Form 483, July 2010

FDA issued this Form 483 to Theken Spine LLC. in Akron, Ohio (FEI 3002498892) at the end of a device inspection that ended July 2, 2010. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jul 2, 2010
Devices · Postmarket Assurance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 17, 2024
4,947 days after the inspection
Facility
FEI 3002498892
Akron, Ohio

FDA's 3 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  2. Individual Report of Malfunction

    An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  3. Documentation

    Corrective and preventive action activities and/or results have not been documented.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.