Skip to content

Encore Laboratory Form 483, January 2026

FDA issued this Form 483 to Encore Laboratory in Santa Ana, California (FEI 3020749939) at the end of a device inspection that ended January 9, 2026. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 23, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jan 9, 2026
Devices · Postmarket Assurance: Devices
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Oct 23, 2026
287 days after the inspection
Facility
FEI 3020749939
Santa Ana, California

FDA's 5 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. DHR content

    The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

  2. Quality Audit/Reaudit - conducted

    Quality audits have not been performed.

  3. Label to bear a unique device identifier

    The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

  4. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  5. Nonconforming product control

    Products that do not conform to specifications are not adequately controlled.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.