Skip to content

American CryoStem Corporation Form 483, July 2017

FDA issued this Form 483 to American CryoStem Corporation in Monmouth Junction, New Jersey (FEI 3008307548) at the end of a biologics inspection that ended July 26, 2017. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jul 26, 2017
Biologics · Human Cellular, Tissue, and Gene Therapies
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Jan 17, 2024
2,366 days after the inspection
Facility
FEI 3008307548
Monmouth Junction, New Jersey

FDA's 9 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Absence of Written Procedures

    There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.

  2. Validation lacking for sterile drug products

    Procedures designed to prevent microbiological contamination of drug products purporting to be sterile do not include validation of the sterilization process.

  3. Lack of written stability program

    There is no written testing program designed to assess the stability characteristics of drug products.

  4. Air filtration system lacking in production area

    The production area air supply lacks an appropriate air filtration system.

  5. Environmental Monitoring System

    Aseptic processing areas are deficient regarding the system for monitoring environmental conditions.

  6. Adequate space lacking to prevent mix-ups and contamination

    The building lacks adequate space for the orderly placement of equipment and materials to prevent mix-ups between different components, in-process materials and drug products and to prevent contamination.

  7. Establish reliability of supplier's C of A

    Establishment of the reliability of the component supplier's report of analyses is deficient in that the test results are not appropriately validated at appropriate intervals.

  8. Procedures To Be in Writing

    Written procedures are lacking which describe in sufficient detail the sampling and testing of components.

  9. Cleaning System

    Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterJan 3, 2018CBER · 161 days after this inspection ended

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.