3B Medical, Inc Form 483, December 2025

FDA issued this Form 483 to 3B Medical, Inc in Dublin, Ohio (FEI 3009096682) at the end of a device inspection that ended December 10, 2025. FDA classified the inspection Official Action Indicated (OAI) and issued the facility a warning letter on May 21, 2026.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 12, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Dec 10, 2025
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Nov 12, 2026
337 days after the inspection
Facility
FEI 3009096682
Dublin, Ohio

FDA's 7 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

  2. Design history file

    The design history file does not demonstrate that the design was developed following the approved design plan.

  3. Design Validation - Risk analysis not performed/inadequate

    Risk analysis is inadequate and is incomplete.

  4. Management review - defined interval, sufficient frequency

    Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

  5. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  6. Training - Lack of or inadequate procedures

    Procedures for training and identifying training needs have not been adequately established.

  7. Report of malfunction likely to cause death or injury

    The importer failed to submit a report to the manufacturer within 30 days concerning information that one of the devices marketed by the importer has malfunctioned and that such device or a similar device marketed by the importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.